Contec CMS800G User Manual Download Page 23

The print output content contains: BED  NO. ④ NAME ④ WEEKS ④ PATIENT NO. ④ FHR1 
trend ④FHR2 trend④Twins Monitoring④④TOCO wave④print speed④date④time.

It also contains other icons like④auto zero mark   ④alarm mark   ④FMOV mark “↑”④event mark 

“↓” etc.

Note(

① 

when paper is used out, printing will stop, and the data will be saved in memory, 

when 

paper reloaded, operator can use wave review function to print the saved data. 

  the monitor has the function of 12-hour wave storage, review and print. Record 

will not saved when monitor is turned off.

 To ensure print precisely, recommend to print and adjust the baseline when paper 

is loaded .

 if the paper coming out from the notch in deflection way when printing, the data 

may be not precise or paper jam will occur, operator should stop printing and 
reload paper.

 Please set all print parameters well before printing, and do not try to change the

 

setup in the process of printing.

5.3 Operation After Monitoring

  Remove  transducers from patient. Wipe  transducer  with a  soft  cloth  to  remove  remaining 

ultrasound coupling gel.

 Tear the paper at the folding place.

 Switch off the power of monitor.

6

Maintenance, Care and Cleaning

To ensure that the monitor works properly, please read the manual and operation procedure as well 
as the maintenance before using the monitor, and operate it as requested.

6.1 Preventive Maintenance

(1) Visual Inspection

The user must check that the equipment, cables and transducers do not have visible evidence of 
damage that may affect patient safety or monitoring capability before use. The recommended 
inspection interval is once per week or less. If damage is evident, replacement is recommended 
before use.

 

   

(2) Routine Inspection

The equipment  should  undergo  periodic  safety  testing  to  insure proper patient  isolation  from 
leakage currents. This should include leakage current measurement and insulation testing. The 
recommended testing interval is once a year or as specified in the institution’s test and inspection  
protocol.

Summary of Contents for CMS800G

Page 1: ...CONTEC MEDICAL SYSTEMS CO LTD CMS800G FETAL MONITOR...

Page 2: ...er and Removing Paper Jam 17 4 5 Power On the Monitor 17 4 6 Connect Transducers 17 5 Monitoring 17 5 1 Operation Procedure 17 5 2 Print Operation 20 5 3 Operation After Monitoring 21 6 Maintenance Ca...

Page 3: ...nstallation of the relevant room complies with national standards and the instrument is used in accordance with the instructions for use Note This device is not intended for home use WARNING This devi...

Page 4: ...sting The recommended testing interval is once per year The protection categories against electric shock of the patient connections are FHR1 FHR2 TOCO MARK This symbol indicates that the instrument is...

Page 5: ...t a power cord with wet hands Make certain that your hands are clean and dry before touching a power cord WARNING The monitor should be installed by an authorized and qualified service engineer WARNIN...

Page 6: ...sing the equipment immediately after moving it from a cold environment to a warm humid location CAUTION Do not immerse transducers in liquid When using solutions use sterile wipes to avoid pouring flu...

Page 7: ...ring of two heart rates for twins This is achieved by using the facilities of two ultrasound transducers and an external contractions TOCO transducer with a recorder The monitor can display FHR TOCO M...

Page 8: ...al broad beam ultrasound transducer is used for monitoring fetal heart rate FHR1 The ultrasound transducer operates at a frequency of 1 0MHz Put the ultrasound transducer on maternal abdomen to transm...

Page 9: ...t Socket for FHR1 TOCO transducer and remote Marker FHR2 socket Socket for FHR2 Transducer Socket for Grounding Cable NET Socket for network Warning Symbol Push LCD Screen rotation lock 3 1 4 Main Int...

Page 10: ...nnel it is I under single fetal monitoring mode which is default I II selectable under twins monitoring mode it can be changed through the main menu 3 Volume level ranging from 0 7 0 stands for sound...

Page 11: ...setting parameters it contains alarm on off status alarm upper limit lower limit alarm postpone time print speed and print time 4 Indicate Bar In this item it includes system time bed No gestational a...

Page 12: ...m when FHR is in alarm range When symbol appears the alarm indicator status is shut off Press the button to enable audio alarm the alarm indicator becomes when FHR is in alarm situation the alarm soun...

Page 13: ...Function Freezing the screen Press the button to stop drawing and the screen becomes in frozen status press the button again to continue drawing This operation will clear the screen 8 Knob Key 1 Press...

Page 14: ...orm Review 5 PRINT SET Print Setup Enter Print Setup 6 MONI TYPE Monitor Mode Optional single twins the default is single fetus 7 LANGUAGE Language Selection Optional Chinese CH English EN the default...

Page 15: ...onds Note 1 When FHR is in alarm situation alarm indicator becomes red 2 When FHR exceeds the alarm limit and time exceed the set alarm delay time continuously alarm will occur and an alarm symbol wil...

Page 16: ...ed patient information 14 SAVE Save Information Save related patient information return to previous menu 15 Exit Return to previous menu Note When adjustable range is letter or number numeral and lett...

Page 17: ...r password under the item USR KEY before click CONFIRM This password is provided by manufacturer or distributor when Contec add new function to upgrade the system Version Revolving the knob key to ent...

Page 18: ...r works properly please read this chapter and Chapter 1 Safety Guidance And follow the steps before using the monitor 4 1 Open the Package and Check Open the package and take out the monitor and acces...

Page 19: ...r cabinet to check for a paper jam then feed the paper again Note Only use the manufacturer approved paper to avoid poor printing quality deflection or paper jam 4 5 Power on the Monitor WARNING If an...

Page 20: ...nsducer to the belt Apply aquasonic coupling gel to the face of the transducer 2 Acquiring the Fetal Heart Signal Determine the location of the fetal heart using palpation or a fetoscope Place the ult...

Page 21: ...the optimum position Positions with strong placental sounds swishing or fetal cord pulse indistinct pulse at fetal rate should be avoided If the fetus is in the cephalic position and the mother is sup...

Page 22: ...al movement At the moment the mark will show in the correspond position of trend wave The count of fetal movement will add 1 after each push of the button Push the button and hold for one second then...

Page 23: ...printing 5 3 Operation After Monitoring Remove transducers from patient Wipe transducer with a soft cloth to remove remaining ultrasound coupling gel Tear the paper at the folding place Switch off the...

Page 24: ...rface of the monitor Use dry and soft cloth to wipe WARNING Unplug the monitor from the AC power source and detach all accessories before cleaning Do not immerse the unit in water or allow liquids to...

Page 25: ...ng solution follow the manufacturer s directions carefully to avoid damaging the transducer 3 Wipe the transducer with a cloth soaked in water 4 Wipe the remained humidity with clean and dried cloth 6...

Page 26: ...tor Note The manufacturer has no responsibility for the effectiveness of controlling infectious disease using these chemical agents Please contact infectious disease experts in your hospital for detai...

Page 27: ...afety in Presence of Flammable Gases not suitable for use in presence of flammable gases Electromagnetic Compatibility Group Class A Mode continuous work Power Supply Working Voltage AC 100V 240V Freq...

Page 28: ...on the operation of pregnant women there will be a mark display in the bottom area of FHR wave display section FHR Alarm Alarm for high and low FHR which exceeds appointed limit A 1 2 Transducers 1 Ul...

Page 29: ...of FHR Symptom Possible Cause Solution Inconstant trace Inconstant display Wrong FHR No The pregnant woman is too fat No Improper ultrasound transducer position Change the position of ultrasound tran...

Page 30: ...rately and neither too tight nor too loose Maternal Movement Relax patient s spirit Fetal Movement Wait for a moment then monitor Too high TOCO sensitivity higher than 100 unit The body pressure from...

Page 31: ...Economic Technical Development Zone Hebei Province PRC Tel 86 335 8015433 Fax 86 335 8015432 E mail cms contecmed com cn Website http www contecmed com cn EC REPRESENTATIVE Shanghai International Hold...

Reviews: