Determine the location of the second fetal signal using palpation or a fetoscope.
Attach the buckle of the ultrasound transducer to the belt. Apply aquasonic coupling gel to the face 
of the transducer. Place the second ultrasound transducer on the abdomen over fetal site and move 
it slowly until the characteristic hoof-beat sound of the fetal heart is heard.
The fetal heart rate value FHR2 will be shown on the screen. At the same time, the ultrasound 
wave will be drawn in yellow color on the screen.

CAUTION: Do not mistake the higher maternal heart rate for fetal heart rate

.

4

Monitor Adjustments:

Adjust the position of ultrasound scanner according to the need.
There is only one fetal heart sound can be heard from the speaker, change it by selecting different 
channel of fetal heart sound (the first sound channel for FHR1, and the second sound channel for  
FHR2)
Readjust the volume setting for the desired loudness.

Note:
The ultrasound transducer measures the FHR;  the misuse of  it  will  be result in wrong 
measurement or misunderstanding of it. So it requires the doctor pay attention to it: 

 The best quality records will only be obtained if the transducer is placed in the optimum 

position.

  Positions with strong placental sounds (swishing) or fetal cord pulse (indistinct pulse at 

fetal rate) should be avoided.

 

If the fetus is in the cephalic position and the mother is supine, the clearest heart sound  

will normally be found on the midline below the umbilicus. During monitoring prolonged 
lying in the supine position should be avoided owing to the possibility of supine hypotension. 
Sitting up or lateral positions are preferable and may be more comfortable to the mother.

 It is not possible to FHR unless an audible fetal heart signal is present. The fetal pulse can 

be   distinguished   from   the   maternal   pulse   by   feeling   the   mother’s   pulse   during   the 
examination.

 During the monitoring, the doctor should observe the monitor screen, if the screen break 

off frequently, the position of the ultrasound transducer may had out of proper position due 
to the moving of the fetus. 

 During the monitoring, if the FHR can be heard without steadily sound of the fetal heart, 

it may not proper positions. So move it slowly until the proper position is found. But if it is  
not found, the doctor should do other examination, to observe if the fetus is normal.

5.1.2 Monitoring Contractions

Operation Procedure:

1

Preparing the Monitor

Turn the monitor on and verify that the normal monitoring screen appears on the display.
Insert the TOCO Transducer into the socket.

Summary of Contents for CMS800G

Page 1: ...CONTEC MEDICAL SYSTEMS CO LTD CMS800G FETAL MONITOR...

Page 2: ...er and Removing Paper Jam 17 4 5 Power On the Monitor 17 4 6 Connect Transducers 17 5 Monitoring 17 5 1 Operation Procedure 17 5 2 Print Operation 20 5 3 Operation After Monitoring 21 6 Maintenance Ca...

Page 3: ...nstallation of the relevant room complies with national standards and the instrument is used in accordance with the instructions for use Note This device is not intended for home use WARNING This devi...

Page 4: ...sting The recommended testing interval is once per year The protection categories against electric shock of the patient connections are FHR1 FHR2 TOCO MARK This symbol indicates that the instrument is...

Page 5: ...t a power cord with wet hands Make certain that your hands are clean and dry before touching a power cord WARNING The monitor should be installed by an authorized and qualified service engineer WARNIN...

Page 6: ...sing the equipment immediately after moving it from a cold environment to a warm humid location CAUTION Do not immerse transducers in liquid When using solutions use sterile wipes to avoid pouring flu...

Page 7: ...ring of two heart rates for twins This is achieved by using the facilities of two ultrasound transducers and an external contractions TOCO transducer with a recorder The monitor can display FHR TOCO M...

Page 8: ...al broad beam ultrasound transducer is used for monitoring fetal heart rate FHR1 The ultrasound transducer operates at a frequency of 1 0MHz Put the ultrasound transducer on maternal abdomen to transm...

Page 9: ...t Socket for FHR1 TOCO transducer and remote Marker FHR2 socket Socket for FHR2 Transducer Socket for Grounding Cable NET Socket for network Warning Symbol Push LCD Screen rotation lock 3 1 4 Main Int...

Page 10: ...nnel it is I under single fetal monitoring mode which is default I II selectable under twins monitoring mode it can be changed through the main menu 3 Volume level ranging from 0 7 0 stands for sound...

Page 11: ...setting parameters it contains alarm on off status alarm upper limit lower limit alarm postpone time print speed and print time 4 Indicate Bar In this item it includes system time bed No gestational a...

Page 12: ...m when FHR is in alarm range When symbol appears the alarm indicator status is shut off Press the button to enable audio alarm the alarm indicator becomes when FHR is in alarm situation the alarm soun...

Page 13: ...Function Freezing the screen Press the button to stop drawing and the screen becomes in frozen status press the button again to continue drawing This operation will clear the screen 8 Knob Key 1 Press...

Page 14: ...orm Review 5 PRINT SET Print Setup Enter Print Setup 6 MONI TYPE Monitor Mode Optional single twins the default is single fetus 7 LANGUAGE Language Selection Optional Chinese CH English EN the default...

Page 15: ...onds Note 1 When FHR is in alarm situation alarm indicator becomes red 2 When FHR exceeds the alarm limit and time exceed the set alarm delay time continuously alarm will occur and an alarm symbol wil...

Page 16: ...ed patient information 14 SAVE Save Information Save related patient information return to previous menu 15 Exit Return to previous menu Note When adjustable range is letter or number numeral and lett...

Page 17: ...r password under the item USR KEY before click CONFIRM This password is provided by manufacturer or distributor when Contec add new function to upgrade the system Version Revolving the knob key to ent...

Page 18: ...r works properly please read this chapter and Chapter 1 Safety Guidance And follow the steps before using the monitor 4 1 Open the Package and Check Open the package and take out the monitor and acces...

Page 19: ...r cabinet to check for a paper jam then feed the paper again Note Only use the manufacturer approved paper to avoid poor printing quality deflection or paper jam 4 5 Power on the Monitor WARNING If an...

Page 20: ...nsducer to the belt Apply aquasonic coupling gel to the face of the transducer 2 Acquiring the Fetal Heart Signal Determine the location of the fetal heart using palpation or a fetoscope Place the ult...

Page 21: ...the optimum position Positions with strong placental sounds swishing or fetal cord pulse indistinct pulse at fetal rate should be avoided If the fetus is in the cephalic position and the mother is sup...

Page 22: ...al movement At the moment the mark will show in the correspond position of trend wave The count of fetal movement will add 1 after each push of the button Push the button and hold for one second then...

Page 23: ...printing 5 3 Operation After Monitoring Remove transducers from patient Wipe transducer with a soft cloth to remove remaining ultrasound coupling gel Tear the paper at the folding place Switch off the...

Page 24: ...rface of the monitor Use dry and soft cloth to wipe WARNING Unplug the monitor from the AC power source and detach all accessories before cleaning Do not immerse the unit in water or allow liquids to...

Page 25: ...ng solution follow the manufacturer s directions carefully to avoid damaging the transducer 3 Wipe the transducer with a cloth soaked in water 4 Wipe the remained humidity with clean and dried cloth 6...

Page 26: ...tor Note The manufacturer has no responsibility for the effectiveness of controlling infectious disease using these chemical agents Please contact infectious disease experts in your hospital for detai...

Page 27: ...afety in Presence of Flammable Gases not suitable for use in presence of flammable gases Electromagnetic Compatibility Group Class A Mode continuous work Power Supply Working Voltage AC 100V 240V Freq...

Page 28: ...on the operation of pregnant women there will be a mark display in the bottom area of FHR wave display section FHR Alarm Alarm for high and low FHR which exceeds appointed limit A 1 2 Transducers 1 Ul...

Page 29: ...of FHR Symptom Possible Cause Solution Inconstant trace Inconstant display Wrong FHR No The pregnant woman is too fat No Improper ultrasound transducer position Change the position of ultrasound tran...

Page 30: ...rately and neither too tight nor too loose Maternal Movement Relax patient s spirit Fetal Movement Wait for a moment then monitor Too high TOCO sensitivity higher than 100 unit The body pressure from...

Page 31: ...Economic Technical Development Zone Hebei Province PRC Tel 86 335 8015433 Fax 86 335 8015432 E mail cms contecmed com cn Website http www contecmed com cn EC REPRESENTATIVE Shanghai International Hold...

Reviews: