Manual Power Supplies
Running condition
Display
600
v
85 mA
51 W
00:29:33 h
1|2 17/12/15 10:36:18
When running an experiment, the display allows to see the actual parameters. It is divided into three
fields:
•
The
upper left
field shows :
◦
a blinking
symbol to indicate there is a possible dangerous condition on the output sockets.
Avoid to contact the outlet sockets and cables!
◦
the selected measured parameter. Use the
éê
keys to switch the parameter in this field.
•
The
upper right
field shows the other 3 parameters.
•
The
bottom
field shows date + time at the right and the actual method/step combination at the
left when such a procedure is running, no indication when running a manual procedure. The
bottom field line is also used for system messages when required.
The unit of the
parameter
that is being kept constant is given with blinking
bold
characters.
Buttons
The keys be used as follows:
•
éê
Buttons for selecting another parameter in the upper left corner of the display .
•
Button to modify the parameters.
•
/
Button to stop or to pause the procedure.
Adjusting parameters
It is possible to modify the parameters while running. These modifications are temporarily and are not
stored in the non volatile memory. It is not possible to modify the settings of a voltage gradient step.
•
Press
:
600
v
1000 mA
150 W
00:29:33 h
éê
=ADJUST,
=NEXT
•
The display shows the preset values for V, mA and W. The timer remains counting down when a
timer value has been set.
•
Choose the limit to adjust with the
button.
•
Adjust the value with the
éê
buttons.
•
The display will return automatically to the normal run mode when no button is pressed for 5
seconds.
Rev. 1605
9