background image

Manual Power Supplies

End of procedure

The procedure will stop automatically when the timer has reached its end value. A signal will 
sound during 10 seconds. The sound signal can be ended by pressing a button.

The procedure can also be stopped by pressing 

/

The power supply will then be in a paused 

condition. It allows to modify the run parameters and/or to view the gel. When pressing 

/

 

again, the power supply will continue the procedure where paused. To stop the procedure 
completely, hold 

 pressed until the main menu appears.

The power supply will remain showing the display of the running condition until the voltage has 

lowered to a safety value.

When the procedure has finished, a report will be shown on the display. Scroll the display lines 
with the  

éê

 buttons to see it completely. Press 

 or 

/

 to return to the main menu.

Alarm messages

In case a problem or a hazardous situation is detected, the power supply will intterupt the experiment 
with one of the following errors. A sound will also be produced that can be stopped by pressing any key. 
The hazardous situations require that the power supply needs to be switched off before starting the 
experiment again.

GROUND LEAKAGE

A dangerous ground leakage is present (check thoroughly the complete assembly). Your power 

supply is 

NOT 

defective! See 

Frequently Asked Questions

 at the end of this manual.

SHORT-CIRCUIT

Connected tank is short-circuited or has a very low resistance (check thoroughly the complete 
assembly). Your power supply is 

NOT 

defective! See 

Frequently Asked Questions

 at the end of 

this manual.

OVERLOAD

Connected tank has a too low resistance (check specifications). Your power supply is 

NOT 

defective! See 

Frequently Asked Questions

 at the end of this manual.

LOW CURRENT

The output current is very low (switch the alarm off in the 

SETUP

 menu). Your power supply is 

NOT 

defective! See 

Frequently Asked Questions

 at the end of this manual.

TECHNICAL PROBLEM

Failure in the electronics. See 

Support

 on 

www.consort.be

. You could also try a general RESET: 

switch the instrument on while holding 

 pressed. Attention! All stored programs and data will 

be erased!

USB Digital Port

This power supply has a digital USB port allowing to interface with a computer. Programs can read and 

control the power supply completely. More information can be found on our support website.

Driver

The driver can be downloaded using the link for the USB drivers on our website: 

http://www.consort.be/downloads/software/

.

                                                                                  

 

                                                                              

Rev. 1605

10

Summary of Contents for EV1450

Page 1: ...MANUAL HANDLEIDING MODE D EMPLOI ANLEITUNG EV1450 EV2150 EV2310 EV2320 EV2650 EV3020 EV3330 EV3610 EV3620 February 2015...

Page 2: ...trast 3 Date Time 4 Password 4 Detection 5 Data 5 Info 5 Method programming 6 Manual programming 8 Running condition 9 Display 9 Buttons 9 Adjusting parameters 9 End of procedure 10 Alarm messages 10...

Page 3: ...sed under normal laboratory conditions and in accordance with the operating limitations and maintenance procedures in this instruction manual and when not having been subject to accident alteration mi...

Page 4: ...safety cover 6 Never make any other connections such as e g putting several power supplies in series or in parallel 7 In order to prevent electric shock never open the back plate nor remove the cover...

Page 5: ...preset the system settings of the power supply for language display contrast real time clock password detections data recording System Language The power supply has several languages built in for the...

Page 6: ...TIME MODIFY 07 12 2015 13 08 47 System Password When desired a password can be set which will be requested when trying to modify the setup or the parameters of a procedure The user can then only run...

Page 7: ...for the remaining time 1 Follow these steps to select LOW CURRENT and press PROCEDURE MANUAL METHOD SETUP SYSTEM DETECTION DATA INFO DETECTION LOW CURRENT POWER FAIL 2 Decide whether or not the output...

Page 8: ...ctrophoresis tank to the instrument 3 Switch the instrument on 4 Select METHOD and press PROCEDURE MANUAL METHOD SETUP 5 Select the desired method and press CHOOSE METHOD 1 6 When previously multiple...

Page 9: ...1 END METHOD NO Choose NO to resume at point 6 with the next step choose YES to stop after this step 14 Press to start the experiment At any time you can return to the main menu by holding pressed At...

Page 10: ...lue and press MANUAL TIMER 00 30h The value format is hh mm when entering a value in hours The format is a value with 2 decimals when entering kiloVolthours Set to zero when required to continue until...

Page 11: ...also used for system messages when required The unit of the parameter that is being kept constant is given with blinking bold characters Buttons The keys be used as follows Buttons for selecting anot...

Page 12: ...e starting the experiment again GROUND LEAKAGE A dangerous ground leakage is present check thoroughly the complete assembly Your power supply is NOT defective See Frequently Asked Questions at the end...

Page 13: ...ep a constant current during a run Program the desired current and a higher voltage and power then the maximum expected values Voltage Current x Resistance Power Voltage x Current How to keep a consta...

Page 14: ...Manual Power Supplies Specifications Rev 1605 12...

Page 15: ...fective equipment returned to it during the warranty period without charge CIF Turnhout prepaid by sender provided the equipment has been used under normal laboratory conditions and in accordance with...

Page 16: ...EMC DIRECTIVE 2004 108 EC EN61326 1 ROHS DIRECTIVE 2011 65 EU EN50581 2012 Consort bvba Parklaan 36 B 2300 Turnhout Belgium Tel 32 0 14 41 12 79 Fax 32 0 14 42 91 79 Name Jan De Ceuster Signature Tit...

Reviews: