End of procedure
•
The procedure will stop automatically when the timer has reached its end value. A signal will
sound during 10 seconds. The sound signal can be ended by pressing a button.
•
The procedure can also be stopped by pressing
/
.
The power supply will then be in a paused
condition. It allows to modify the run parameters and/or to view the gel. When pressing
/
again, the power supply will continue the procedure where paused. To stop the procedure
completely, hold
pressed until the main menu appears.
•
The power supply will remain showing the display of the running condition until the voltage has
lowered to a safety value.
•
When the procedure has finished, a report will be shown on the display. Scroll the display lines
with the
éê
buttons to see it completely. Press
or
/
to return to the main menu.
Alarm messages
In case a problem or a hazardous situation is detected, the power supply will intterupt the experiment
with one of the following errors. A sound will also be produced that can be stopped by pressing any key.
The hazardous situations require that the power supply needs to be switched off before starting the
experiment again.
•
GROUND LEAKAGE
A dangerous ground leakage is present (check thoroughly the complete assembly). Your power
supply is
NOT
defective! See
Frequently Asked Questions
at the end of this manual.
•
SHORT-CIRCUIT
Connected tank is short-circuited or has a very low resistance (check thoroughly the complete
assembly). Your power supply is
NOT
defective! See
Frequently Asked Questions
at the end of
this manual.
•
OVERLOAD
Connected tank has a too low resistance (check specifications). Your power supply is
NOT
defective! See
Frequently Asked Questions
at the end of this manual.
•
LOW CURRENT
The output current is very low (switch the alarm off in the
SETUP
menu). Your power supply is
NOT
defective! See
Frequently Asked Questions
at the end of this manual.
•
TECHNICAL PROBLEM
Failure in the electronics. See
Support
on
www.consort.be
. You could also try a general RESET:
switch the instrument on while holding
pressed. Attention! All stored programs and data will
be erased!
USB Digital Port
This power supply has a digital USB port allowing to interface with a computer. Programs can read and
control the power supply completely. More information can be found on our support website.
Driver
The driver can be downloaded using the link for the USB drivers on our website:
http://www.consort.be/downloads/software/
.