background image

88

E3xxx

E3xxx

Gel staining and viewing

Gel staining and viewing

All trays and the E3100 unit allow staining to be performed with-

out removing the gel from the tray if this is preferred.

1

. Transfer the gel to a vessel containing  the appropriate vol-

ume of 0.5 µg/ml ethidium bromide stain for 15...30 minutes, 

see solutions for stock stain concentration and adjust to the 

volume used accordingly. The entire gel should be covered. 

(Caution: ethidium bromide is a suspected carcinogen and the 

necessary safety precautions should be undertaken).

2

. De-stain  the  gel  for  10–30  minutes  in  distilled  water  again 

ensuring the gel is completely immersed. 

3

. Rinse the gel twice for a couple of seconds with distilled wa-

ter.

4

. Transfer the gel to a UV Transilluminator.

5

. The samples will often appear as brighter, clearer bands when 

photographed or viewed using a gel documentation system. 

However if the gel bands are too faint then the staining pro-

cedure should be adjusted so that there is less de-staining. 

If there is too much background then the staining procedure 

should be adjusted so that there is more de-staining. 

Running the gel

Running the gel

1

. Mix the sample to be loaded with sample buffer (see solutions 

for common sample buffers). Usually 3 µl of sample buffer is 

adequate but less may be used with sample volumes of less 

than 10 µl.

2

. Fill the unit with buffer until the gel is just flooded with buff-

er. This will give the fastest resolution times. For enhanced 

quality of resolution of sample, fill the unit to 5 mm above 

the gel. 

3

. Load the samples into the wells using pipettes. Multi-channel 

pipettes  can  be  used  for  loading  samples  with  MultiPipette 

compatible combs, see listing in accessories for identification 

of these.

4

. Carefully place the lid on the tank and connect to a power 

supply.

5

. Typically  gels  are  run  at  between  90  and  150  V.  However, 

maximum  voltage  is  indicated  on  the  serial  badge  of  each 

unit. It should be noted that a higher voltage generally give 

faster but poorer quality sample resolution.

Summary of Contents for E3100

Page 1: ...Consort MANUAL E3100 E3200 E3300 E3400 E3500 E3600 July 2009 Distributor Topac Inc 231 CJC Highway Cohasset MA 02025 Tel 781 740 8778 Fax 781 740 8779 www topac com email sales topac com ...

Page 2: ...tal conditions 3 Fitting electrode cables 4 Fitting loading guides 4 Gel preparation 5 Gel pouring with casting dams 6 Gel pouring with model E3100 6 Gel pouring using tape method 7 Gel pouring using the flexicaster 7 Running the gel 8 Gel staining and viewing 8 Solutions 9 Troubleshooting 10 ...

Page 3: ...d maintenance procedures out lined in this instruction manual and when not having been subject to accident alteration misuse or abuse No liability is accepted for loss or damage arising from the incor rect use of this unit CONSORT s liability is limited to the repair or replacement of the unit or refund of the purchase price at CONSORT s option CONSORT is not liable for any consequential damages C...

Page 4: ...connect the leads Do not exceed the maximum operating voltage or current Do not operate the electrophoresis units in metal trays Acrylamide is a volatile cumulative neurotoxin and suspect ed carcinogen Wear effective protective clothing and follow recommended handling and disposal procedures Polymerised gels contain some unpolymerised monomer Han dle with gloves only Following the replacement of a...

Page 5: ...re should be taken not to damage the enclosed electrode and vigorous cleaning is not necessary or advised Air drying is preferably before use The units should only be cleaned with the following warm water with a mild concentration of soap or other mild de tergent compatible detergents include dish washing liquid hexane and aliphatic hydrocarbons The units should not be left to in detergents for mo...

Page 6: ...p between the lid and the leading edge of no gap between the lid and the leading edge of the cable fitting 4 Refit the lid Fitting loading guides Fitting loading guides These can be fitted to enhance visibility of the wells if desired They can be fitted to the white vinyl platform sheet or to the unit itself 1 Seat the tray in the unit and note the position of the comb grooves The samples run blac...

Page 7: ...ical flask 3 Add the appropriate amount of 1 x TAE or TBE solution from the table above To prevent evaporation during the dissolving steps below the conical flask should be covered with para film 4 Dissolve the agarose powder by heating the agarose either on a magnetic hot plate with stirring bar or in a microwave oven If using the microwave method the microwave should be set at around a 400 W or ...

Page 8: ...tank the gel including tray to the main tank Gel pouring Gel pouring with model E3100 with model E3100 Gel pouring with model E3100 Gel pouring Gel pouring with model E3100 Gel pouring 1 Fit the casting gates into the grooves in the unit which are 10 mm from the platinum wire Ensure that these are seated as deeply as possible This will ensure that a good seal is formed and that there is no possibi...

Page 9: ...agarose to set ensuring that the gel remains undis turbed 5 Carefully remove the gel casting gates and comb and transfer the gel including tray to the main tank the gel including tray to the main tank Gel pouring using the flexicaster Gel pouring using the flexicaster 1 Level the flexicaster base by adjusting the feet so that the bubble is exactly central 2 Insert the desired length tray into the ...

Page 10: ... faint then the staining pro cedure should be adjusted so that there is less de staining If there is too much background then the staining procedure should be adjusted so that there is more de staining Running the gel Running the gel 1 Mix the sample to be loaded with sample buffer see solutions for common sample buffers Usually 3 µl of sample buffer is adequate but less may be used with sample vo...

Page 11: ... tris pH 7 6 89 mM boric acid 2 mM EDTA 10 x 1l dissolve in 750 ml distilled water 108 g tris base FW 121 55 g boric acid FW 61 8 40 ml 0 5 M EDTA pH 8 0 Fill to 1 litre with distilled water Sample Loading Dye 10x sample buffer stock consists of 50 glycerol 0 25 bromophenol blue and 0 25 xylene cyanole FF in 1x TAE buffer Only 1 10 ml of the 10x loading dye should be pre pared Ethidium Bromide Sol...

Page 12: ...nt gel from breaking Avoid damaging the well with the pipette when loading the sample Aim for the centre of the well and avoid damaging the bottom of the well with the pipette tip Samples leak underneath the gel upon loading The bottom of the wells were torn when the comb was re moved To avoid this tearing carefully wiggle the comb to free the teeth from the gel Distorted sample wells Incomplete p...

Page 13: ...erol in the sample Poor band resolution Add ficoll glycerol or sucrose to the sample loading buffer to ensure that the sample layers on the bottom of the well Ensure sample is completely dissolved Reduce voltage sample concentration or sample volume Ensure there is at least 1 mm of gel below the bottom of the comb to prevent samples from leaking out the bottom of the well Reduce salt concentration...

Page 14: ...12 E3xxx E3xxx ...

Page 15: ... E3100 E3200 E3300 E3400 E3500 E3600 to which this declaration relates is in conformity with the following standards EN61010 LOW VOLTAGE DIRECTIVE 73 23 EEG EN50081 1 EN50082 1 EN60555 2 EMC DIRECTIVE 89 336 EEG Turnhout June 25 2009 CONSORT nv Tel 32 14 41 12 79 Parklaan 36 Fax 32 14 42 91 79 B 2300 Turnhout E mail info consort be ...

Page 16: ......

Reviews: