19

acoustic or visual test (display of measurement) is possible. An acou-
stic signal is emitted at resistors below 40 Ohm.

c) Negative Polarity

Shows if test-leads are confused or indicates negative input with a “-“
before the measurement.

d)  FREQ Frequency Measurement

You can measure frequency up to 40 MHz.

e)  TEMP Temperature Measurement

You can measure temperatures from -40°C to +1200°C with a ther-
moelement (NiCrNi).

f) CAP Capacitance Measurement

The capacitance range lets you measure discharged capacitors of 4 nF
to 400 µF.

g)  HFE Transistor Test

Enables you to measure the hFE value.

h)  LOGIC Logictest

This function enables you to check and to display all usual logic levels.

i) 

Low Battery 

A 9-Volt alkaline battery in this meter has an average life of approx.
20 to 30 hours. Low battery indication appears approx. 8 hours befo-
re battery is “dead“.

j)  The following symbols indicate the units of the value displays:

AC

=  AC current or voltage

DC

=  DC current or voltage

mV

=  Millivolt (exp.-3)

V

=  Volts

mA

=  Milliampere (exp.-3)

A

= Ampere

W

=  Watt

kWh

=  Kilowatt hours

Hz

=  Hertz

KHz

=  Kilohertz (exp.3)

MHz

=  Megahertz (exp.6)

+

-

18

recalled value (ref. no. 1) on the middle sub-display and 3rd recalled
value (ref. no. 2) on the right sub-display.

i) CMP (= Comparison)

This function mode enables you to make the High/Low comparison of
subsequent readings, by comparing a reading with both the stored
high reference value and the stored low reference value.

On the sub-display “Lo“ appears for reading less than the low refe-
rence value, “Hi“ is displayed for reading more than the highest refe-
rence value and “Pass“ for a middle value between the low and the
high reference value.
To activate this function mode, push “Func“ until the symbol “CMP“
appears on the top of the display and “LOW“ flashes over the left
sub-display. Now push again SET/R button and then “UP“ and
“DOWN“ to set the minimum reference value with polarity. Press
after each change SET/R button again, to enter the setting and to go
to the next digit. As soon as the minimum reference (MIN) value has
been fixed, “HIGH“ flashes over the right sub-display. Set this high
reference value respectively (with polarity and four digit value). After
the reference values have been defined press SET/R key again. Now
“R-H“ for Range Hold flashes in the top line of the display. Set with
the UP and DOWN buttons the measuring range (e.g. voltage control
in a circuit under worst-case conditions). After the measuring range
has been entered (as far as a manual selection is possible) Comparison
= CMP is activated.

To exit CMP mode turn the rotary function switch (observe the safety
rules!) to an adjacent range or press “FUNC“.
Reset with the SET/R button is also possible, even during measurement.

4.4.5 Display Annunciators and Symbols for the Modes

a) Diode Test  

The value displayed is the forward voltage at approx. 1 mA test cur-
rent. Range of 0 - 2.0 V.

b) Continuity Check 

Enables you to check continuity of wiring, connections or fuses. Either

Summary of Contents for M 3860-M

Page 1: ...is unit left the factory in safe and perfect condition To main tain the unit in this condition and to guarantee a safe operation the user must observe the safety rules and warnings which are contained in this manual by all means This multimeter may only be used in fuse lines which are protected with 16 A The voltage existing appearing must not exceed 250 VDC VAC and the maximum load must not excee...

Page 2: ... protected against arcing IEC 1010 2 031 section 13 101 Keep children away from measuring units Pay attention to the rules for prevention of accidents in industrial enterprises prescibed by the Industrial Trade Associations for Electrical Installations and Production Facilities When using the unit in schools and hobby workshops the usage and the measurement has to be controlled by the responsible ...

Page 3: ...xceeded overflow no acoustic signal during resistance measure ment diode test or temperature measurement 6 If there are doubts whether a safe usage is still possible the unit has to be put out of operation and be secured against unintentional use It must be assumed that a safe usage is not possible if the unit shows visible damage the unit does not work and longterm storage under unfavourable cond...

Page 4: ...ompartment Now lift the cover carefully use a small screw driver Separate the old battery from the battery snap and fasten a new one of the same type Do not forget to use the provided battery insulation capsule After the battery has been changed reinsert the battery into the bat tery case and close it carefully secure it with the screw 8 Attention Observe the max input limits 19 Data Hold D H Data...

Page 5: ...L CMP c SET R Press this button once to activate or to enter the selected function mode Press in the function mode the button D H again to exit wait for beeper sound 10 Attention Never use the measuring unit before the cover has been closed comple tely to avoid an electrical shock Never leave empty batteries in the measuring unit as even corrosion free batteries might leak and chemicals could be r...

Page 6: ...yed After the first push of the SET R key the A H symbol is set and will stop blin king The R H symbol continues flashing to be able to set the manual measuring range with buttons UP and DOWN If the SET R key is pressed again the R H symbol is fixed You can commence measurement Another third push of the SET R key discards the function and the previous normal mode is restored In the function modes ...

Page 7: ...ode you can freeze a reading in the second small display by pressing the SET R button d Range Hold R H In this function you can switch off the automatic range selection Auto Range and set the measuring range manually Each time the UP or DOWN button is pressed the decimal point is moved one posi tion to the left or to the right the units of measure are changed res pectively e g from kHz to MHz duri...

Page 8: ...y 2nd 16 return from this function mode press Set Reset key twice RESET and you will return to Auto Range e Auto Hold A H In this function the meter will automatically record the minimum and the maximum display reading MIN and MAX values These values are currently updated refreshed The MIN value is shown on the left sub display the MAX value on the right sub display On the middle sub display the a...

Page 9: ...eference value and the stored low reference value On the sub display Lo appears for reading less than the low refe rence value Hi is displayed for reading more than the highest refe rence value and Pass for a middle value between the low and the high reference value To activate this function mode push Func until the symbol CMP appears on the top of the display and LOW flashes over the left sub dis...

Page 10: ...lay the measuring value appears 2 seconds later and in the right sub display the measuring value appears 3 seconds later This means the left sub display shows the measurement value which was 1 second earlier on the main display the middle sub display shows the reading which was 2 seconds earlier in the main display and the right sub display shows the value which was 3 seconds earlier on the main d...

Page 11: ... the value which was 3 seconds earlier in the main display 5 3 Continuity Test With this function you can test dead leads fuses circuits Proceed as follows Set the rotary function switch to Ohm and press DC Ohm AC button until the symbol is displayed Afterwards connect the test lead tips A Current Source Object Multimeter 22 AC Voltage Measurement DC Voltage Measurement 5 2 Current Measurement Fol...

Page 12: ...ds to stabilize If the measured resistance value exceeds the range OL appears in the display or the line is interrupted high ohmic 40 MOhm Hint During resistance measurement the three sub displays below the main display are active In the left sub display the present reading is taken 1 second later the middle display the reading is taken 2 seconds later and in the right display the reading appears ...

Page 13: ...dle sub display shows the reading which was 2 seconds earlier in the main display and the right sub display shows the value which was 3 seconds earlier on the main display 5 5 Diode Test Set the measuring function switch 10 to The annunciator for diodes is displayed Manual range selection is not possible Now connect the test lead tips with the object under measurement an off circuit semiconductor ...

Page 14: ...n the main display the middle sub display shows the reading which was 2 seconds earlier in the main display and the right sub display shows the value which was 3 seconds earlier in the main display 5 8 Transistor Test Attention The transistor jack is not protected against overload You will void war ranty and you can damage the meter if you build and use your own external transistor socket and conn...

Page 15: ...to hFE position 2 Insert the transistor you want to measure into the appropriate transi stor socket Pay attention to the following points observe the sequence of connection of the transistor you will find it in the transistor comparative table Some Darlingtons or special power transistors contain internal base to emitter resistance that could cause undependable measu ring results The hFE measuring...

Page 16: ...troyed Attention Life Danger 32 5 10 Usage of the Analogous Bargraph The bargraph is easy to use and to understand Is is comparable with the needle of an analogous measuring instrument but without its mechani cal disadvantages Is is especially for quick changing measuring signals for which the digital display is too slow So you can quickly recognize see tendencies in the measuring values If the ra...

Page 17: ...s displayed power factor Cos phi is the power factor between Ampere and Voltage it is always 1 if ohmic loads consumers are concerned For inductive consumers cos Phi is between 0 3 and 0 98 The main display will show the true watts of your consumer 4 In the second function mode push UP button once the reference no 1 appears in mode digit on the left The left sub display will show costs per 1 hour ...

Page 18: ... displayed in the left sub display the main display will show the respective frequency 0 010 KHz 10 Hz In the middle sub display the external DC input voltage is displayed in V max 40 V 5 To select other output frequencies push either UP or DOWN button For frequency and connected reference number see below 0 010 KHZ 10 Hz reference no 0 0 050 KHz 50 Hz reference no 1 0 060 KHz 60 Hz reference no 2...

Page 19: ...eading rdg number of digit s dgt s Accuracy is specified at 23 C 5 K with a relative humidity of below 75 not wetting for a period of 1 year after production Warm up time is 1 minute 38 BYTE 1 2 3 4 5 6 7 8 9 A B C D E Example 1 DC 3 9 9 9 V CR Example 2 3 9 9 9 MOh m CR The sequence of data transmission from the meter Main display k left sub display k middle sub display k right sub display k main...

Page 20: ...0 µF 10 nF 400 µF 100 nF Diode Test range 0 to 2000 mV 1 mV test current 1 5 mA max at 1 KOhm forward voltage max 2 0 V 40 Mode Measuring Range Accuracy Resolution DC 400 mV 0 3 1dgt 100 µV Voltage 4 V 1 mV 40 V 10 mV 400 V 100 mV 1000 V 0 5 1dgt 1 V AC 400 mV 0 8 3dgts 100 µV Voltage 4 V 1 mV 40 V 10 mV 400 V 100 mV 750 V 0 8 3dgts 1 V valid at a frequency of 40 Hz to 10 KHz DC 400 µA 1 0 1dgt 0 ...

Page 21: ... 42 7 3 Max Input Limits Voltage Measurement 1000 VDC or 750 VACrms Current Measurement 20 A DC AC A range 400 mA AC DC in the mA range Mode MeasuringRange Accuracy Resolution Frequency 4 KHz 1 0 1dgt 1 Hz 40 KHz 10 Hz 400 KHz 100 Hz 4 MHz 1 KHz 40 MHz 10 KHz Sensitivity 1 Vrms Temperatur 40 C to 200 C 3 5dgts 1 C 200 C to 1200 C 3 2dgts 1 C Power Measurement AC Current max 16 A 2 5dgts AC Voltage...

Reviews: