17
Minimum recommended sterilization exposure times are as follows:
Table 1: Sterilization Parameters
Sterilization Type
Minimum
Temperature
Recommended
Minimum
Exposure Time
Dry Time
Sterilization Parameters for sterilizing the PowerPro Pneumatic Handpieces
(PRO6150, PRO6175, PRO6185), Associated Attachments, and Hose, with or without the
PowerPro PRO5090 Sterilization Case
Steam Pre-vacuum
270°F (132°C)
4 minutes
8 minutes minimum
Steam Gravity
270°F (132°C)
10 minutes
8 minutes minimum
Steam Gravity
250°F (121°C)
40 minutes
8 minutes minimum
CAUTION: The recommended dry cycle must be run on all handpieces and attachments
every time the product is sterilized. Failure to use a dry cycle on the products may lead to
reduced product performance or premature product failure. Operation of a handpiece that is
not completely cool and dry may decrease performance and/or reliability.