Step7- 

It is now time to pick up the test weight and raise to the weighing height 

, (Raise to 1 , Lower to 2) 

 

. Allow approx 2-3 seconds for the weight to settle, 

and then press the 

“Set Laden Calibration” key once.  

 
Step8- 

Check the “Calculated Current Weight” screen to see if the weight 

displays the test weight figure. If not, you may need to reset the Zero, Step 5, 
and the Test Weight again.

   

 
Step 9 

– Press the “Exit” button one press at a time to return to the main 

weighing page and to save the calibration.  
 
Perform as many test weighings until you are satisfied the system is working 
correctly.  
 
If at any time the indicator indicates an error message during the calibration 
process, call your supplier for assistance. Before calling check that the 
hydraulics have been installed into the correct ‘LIFT’ circuit of the machine, 
and also check that your test weight is at least 1/3 - ½ machine capacity.

 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for CL2000MKII

Page 1: ...N and OPERATING INSTRUCTIONS for the COMPULOAD CL2000MKII Series LOAD WEIGHING SYSTEM For forklift or small loader application Instant Weight Pty Ltd P O Box 2340 MIDLAND WA 6936 Tel 08 9274 8600 Emai...

Page 2: ...2 INDEX Introduction Installation Instructions Basic Calibration Basic Weighing WARRANTY...

Page 3: ...klifts and small wheel loaders Operating Voltage is 12 24Volts DC The CL2000MKII will generally be installed with a Hydraulic pressure transducer mounted in the Lift Circuit of the machine hydraulics...

Page 4: ...cessary FITTING THE COMPULOAD TO THE OPERATORS CABIN Select a suitable location and using the mounting bracket as a template drill two holes and mount the bracket into position Allow room for the cabl...

Page 5: ...ff point Failing this the system will have to be entered by direct drilling and tapping into a suitable hydraulic connection Ensure the suitable hydraulic connection tee piece elbow or similar is remo...

Page 6: ...6 Transducer connector instructional video Transducer Connector Diagram Transducer Green...

Page 7: ...rown Power White Earth Green TXD ESTABLISHING A WEIGHING POINT Static mode When weighing in the static method a predetermined weighing point must be established The weighing point on a standard forkli...

Page 8: ...When lowering to position 2 this must be done smoothly and slowly to allow the hydraulic pressure to be consistent Lowering abruptly will cause a pressure spike within the lift circuit and readings wi...

Page 9: ...d power up and go through a short power up cycle Once the power up cycle finishes the indicator will show a random reading in the Current Load screen To ensure you have the hydraulic pressure transduc...

Page 10: ...he Zero Step 5 and the Test Weight again Step 9 Press the Exit button one press at a time to return to the main weighing page and to save the calibration Perform as many test weighings until you are s...

Page 11: ...se to 1 Lower to 2 and on the carriage Arrow or other markings on a loader 1 With the forks empty and mast upright raise forks up to the weighing position If other than 0 00 zero is displayed press ZE...

Page 12: ...te further buckets 4 To CLEAR the TOTAL press the CLEAR button and HOLD for 3 seconds The TOTAL will return to zero to allow the next Load to be completed 5 NOTE pushing the CLEAR button once will del...

Page 13: ...Date each bucket load and bucket number When the CLEAR button is pressed for 3 seconds at the end of a load the docket will finalise and print out a Total 2 PRINTER OFF Whilst loading the printer will...

Page 14: ...print most likely you have the paper upside down Thermal Paper only prints on one side DO NOT PULL THE PAPER THROUGH THE PRINTER THIS WILL STRIP THE GEARS OF THE PRINTER AND CAUSE DAMAGE THAT CANNOT...

Page 15: ...ns are carried out or attempted to be carried out by other than ATLAS WEIGHING Pty Ltd The warranty is not transferable and applies only to the original purchaser unless authorised by ATLAS WEIGHING P...

Reviews: