Section 3 

 

IEC 60601-2-27 Option 

 
This section describes how to set up and use the 
Defib-5 with the IEC 60601-2-27 option.. This 
option allows the Defib-5 to act as the circuits 
described in IEC-60601, Figures 101, 104 and 
105.  Please refer to Figures 3, 4 and 5 for a 
schematic of the circuits in IEC 60601, and 
Figure 1-1 for the appropriate front panel 
connections needed to conduct these tests with 
the Defib-5. 
The connection information for the Defib-5 is 
shown in blue text on the following Figures.   
 
  
 

Notes on testing with the IEC 
60601-2-27 option 

 
Testing with the IEC 60601-2-17 option requires 
minor additions to the circuitry in the Defib-5, 
and these changes are accomplished using 
jumpers on the front panel. 
 
The changes to the circuit are as follows, along 
with the appropriate actions to perform on the 
front panel: 
 
1.

 

50 OHM jack:  This is the standard output 
jack for IEC 60601-1 Table 50, and it is also 
used for the pulse output for testing in IEC 
60601-2-27.   This output provides the 50 

ohm output impedance required by IEC 
60601-2-27. 

 
2.

 

Impedance Jumper:  For testing to IEC 
60601, a 500uH inductance is required.  To 
provide this inductance, the front panel 
jumper must be used to SHORT the two 
impedance jacks. 

 
3.

 

EC-13 SENSE jack:  This jack allows 
verification of the 100 ohm resistor used in 
the Defib-5.  The 100 ohm resistance value 
is measured between the EC-13 SENSE jack 
and the RETURN jack on the front panel of 
the Defib-5.   

 
4.

 

MEAS. POINT:  This is the point which is 
connected to the voltage divider network 
within the Defib-5 to determine PASS/FAIL 
results of the test.  See Fig. 3-1 for 
schematic details. 

 
5.

 

Y1 – Y2:  These differential outputs are to 
be connected to two different channels of an 
oscilloscope, then the outputs are subtracted.  
This resulting waveform is used to judge 
acceptability of the EUT.   

 
 
 

Summary of Contents for TEquipment Defib-5

Page 1: ...Defib 5 Addendum Defibrillation Tester Options Instruction Manual COMPLIANCE WESTUSA...

Page 2: ...t you with your Defib 5 This manual describes options you have purchased with the Defib 5 The operating instructions for the options are included in the Manual so please retain this information for fu...

Page 3: ...Contents of this Manual 1 EC 13 EN 60601 2 49 Option 3 Notes on testing with the EC 13 option 3 IEC 60601 2 27 Option 5 Notes on testing with the IEC 60601 2 27 option 5 Defib 5 Pass Fail Reference 9...

Page 4: ...escribed in EC 13 Figures 9A and 9B EN 60601 2 49 Figure 101 with EC 13 option and EN 60601 2 27 Figures 101 104 and 105 In this section we show the various outputs on the front panel of the Defib 5 a...

Page 5: ...2 Figure 1 MegaPulse Front Panel Outputs...

Page 6: ...ont panel and to Figure 2 1 for the schematic location The 100 ohm resistor value is measured between the EC 13 SENSE jack and the RETURN jack on the front panel 2 Inductance Jumper For testing to EC...

Page 7: ...4 Figure 2 EC 13 Figures 6 and 6A...

Page 8: ...n the front panel 1 50 OHM jack This is the standard output jack for IEC 60601 1 Table 50 and it is also used for the pulse output for testing in IEC 60601 2 27 This output provides the 50 ohm output...

Page 9: ...6 Figure 3 IEC 60601 2 27 Fig 101...

Page 10: ...7 Figure 4 IEC 60601 2 27 Figure 104...

Page 11: ...8 Figure 5 IEC 60601 2 27 Figure 105...

Page 12: ...eference Connect the MEAS POINT jack to the BLUE electrode on the Pass Fail Reference There is no ground connection on the Pass Fail reference which takes the place of the DUT in EN 60601 Figure 50 Os...

Page 13: ...10...

Page 14: ...6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 5 00 PM Pacific Time Compliance West USA 2120 Jimmy Durante Blvd Suite 124 Del Mar CA 92014 Phone 858 4...

Reviews: