EC-13/ EN 60601-2-49 Option 

 
This option allows the Defib-5 to act as the 
circuit described in EC-13, Figures 9 and 9A or 
EN 60601-2-49, Figure 101.  Please refer to 
Figure 2-1 for a schematic of the circuit in EC-
13, Figures 9 and 9A, and Figure 1 for the 
appropriate front panel connections needed to 
conduct these tests with the Defib-5.  Test 
connections are the same for EN 60601-2-49, 
Figure 101. 
The connection information for the Defib-5 is 
shown in green text on Figure 2.   
 
 

Notes on testing with the EC-13 
option 

 
Testing with the EC-13 option requires 
substantial changes to the circuitry in the Defib-
5, and these changes are accomplished using 
jumpers on the front panel, and by using 
different output jacks from the EN 60601 
outputs.  Some models may have jacks on the 
rear panel instead of the front panel.  The front-
panel jack representation is shown in Fig. 1 for 
clarity. 
 
The changes to the circuit are as follows, along 
with the appropriate actions to perform on the 
front panel: 
 
1.

 

EC 13 SENSE Jack:  This jack is necessary 
to verify the value of the 100 ohm resistor.  
(Refer to Figure 1-1 for the jack location on 
the front panel, and to Figure 2-1 for the 
schematic location.)  The 100 ohm resistor 
value is measured between the EC 13 
SENSE  jack and the RETURN jack on the 
front panel. 

 
2.

 

Inductance Jumper:  For testing to EC-13, a 
25uH inductance is required.  To provide 
this inductance, the provided front panel 
jumper must be REMOVED. 

 
3.

 

EC-13 S2:  This switch is provided to 
protect the Signal Generator from the output 
pulse of the Defib-5 when CLOSED, and to 
allow the required 5V p-p signal to be 
injected when OPEN. 

 
4.

 

10 HZ, 5V P-P INPUT:  BNC provided on 
front panel is for connection to the signal 
generator required for the Defibrillator 
Overload Test as shown in EC-13, Fig. 9A. 

 
5.

 

400 OHM Jack:  This is the output jack 
from the Defib-5 used with EC-13 Testing.  
It provides the correct pulse to the EUT, in 
accordance with EC-13 Figures 9A and 9B.  
Do not use the 50 OHM output jack, which 
is provided for testing to EN 60601. 

 
6.

 

RETURN Jack:  This jack is used for the 
return of the test voltage, named P2 in EC-
13. 

 
7.

 

MEAS. POINT:  This is the point which is 
connected to the voltage divider network to 
determine PASS/FAIL results of the test.  
See Fig. 2 for more details. 

 
8.

 

Y1 – Y2:  These differential outputs are to 
be connected to two different channels of an 
oscilloscope, then the outputs are subtracted.  
This resulting waveform is used to judge 
acceptability of the EUT. 

 

 

Summary of Contents for TEquipment Defib-5

Page 1: ...Defib 5 Addendum Defibrillation Tester Options Instruction Manual COMPLIANCE WESTUSA...

Page 2: ...t you with your Defib 5 This manual describes options you have purchased with the Defib 5 The operating instructions for the options are included in the Manual so please retain this information for fu...

Page 3: ...Contents of this Manual 1 EC 13 EN 60601 2 49 Option 3 Notes on testing with the EC 13 option 3 IEC 60601 2 27 Option 5 Notes on testing with the IEC 60601 2 27 option 5 Defib 5 Pass Fail Reference 9...

Page 4: ...escribed in EC 13 Figures 9A and 9B EN 60601 2 49 Figure 101 with EC 13 option and EN 60601 2 27 Figures 101 104 and 105 In this section we show the various outputs on the front panel of the Defib 5 a...

Page 5: ...2 Figure 1 MegaPulse Front Panel Outputs...

Page 6: ...ont panel and to Figure 2 1 for the schematic location The 100 ohm resistor value is measured between the EC 13 SENSE jack and the RETURN jack on the front panel 2 Inductance Jumper For testing to EC...

Page 7: ...4 Figure 2 EC 13 Figures 6 and 6A...

Page 8: ...n the front panel 1 50 OHM jack This is the standard output jack for IEC 60601 1 Table 50 and it is also used for the pulse output for testing in IEC 60601 2 27 This output provides the 50 ohm output...

Page 9: ...6 Figure 3 IEC 60601 2 27 Fig 101...

Page 10: ...7 Figure 4 IEC 60601 2 27 Figure 104...

Page 11: ...8 Figure 5 IEC 60601 2 27 Figure 105...

Page 12: ...eference Connect the MEAS POINT jack to the BLUE electrode on the Pass Fail Reference There is no ground connection on the Pass Fail reference which takes the place of the DUT in EN 60601 Figure 50 Os...

Page 13: ...10...

Page 14: ...6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 5 00 PM Pacific Time Compliance West USA 2120 Jimmy Durante Blvd Suite 124 Del Mar CA 92014 Phone 858 4...

Reviews: