Defib-5 Pass/Fail Reference 

 

 

The Defib-5 Pass/Fail Reference is designed to 
allow the operator to ensure operation  of the 
Defib-5 high voltage output circuit and the 
measurement circuit at full voltage.  It augments 
Section “Verification of Measurement Circuit 

Operation” of the MegaPulse Instruction 
Manual. 
 

Description 

 
The Defib-5 Pass/Fail Reference relies on a 
common ground between the Return jack and the 
oscilloscope. 

 
To use: 
 
Connect the OUTPUT (or 50 OHM) jack to the 
RED electrode on the Pass/Fail Reference. 
 
Connect the MEAS. POINT jack to the BLUE 
electrode on the Pass/Fail Reference. 
 
There is no ground connection on the Pass/Fail 
reference, which takes the place of the DUT in 
EN 60601, Figure 50.  Oscilloscope ground is 
supplied by the BNC jack connections. 
 
Connect the Y1 and Y2 outputs of the Defib-5 to 
an oscilloscope in accordance with the 

instructions in the MegaPulse Instruction 
Manual, Defib-5 section. 
 
Set the oscilloscope in accordance with the 
instructions in Paragraph 1 of the Verification of 
Measurement Circuit Operation. 
 
Set the Defib-5 meter to read 5000V, and 
Trigger the output. 
 
The oscilloscope should display a waveform 
similar to Figure D-2 in the MegaPulse Manual, 
with a peak value of 1 volt. 
 
 
 

 

Summary of Contents for TEquipment Defib-5

Page 1: ...Defib 5 Addendum Defibrillation Tester Options Instruction Manual COMPLIANCE WESTUSA...

Page 2: ...t you with your Defib 5 This manual describes options you have purchased with the Defib 5 The operating instructions for the options are included in the Manual so please retain this information for fu...

Page 3: ...Contents of this Manual 1 EC 13 EN 60601 2 49 Option 3 Notes on testing with the EC 13 option 3 IEC 60601 2 27 Option 5 Notes on testing with the IEC 60601 2 27 option 5 Defib 5 Pass Fail Reference 9...

Page 4: ...escribed in EC 13 Figures 9A and 9B EN 60601 2 49 Figure 101 with EC 13 option and EN 60601 2 27 Figures 101 104 and 105 In this section we show the various outputs on the front panel of the Defib 5 a...

Page 5: ...2 Figure 1 MegaPulse Front Panel Outputs...

Page 6: ...ont panel and to Figure 2 1 for the schematic location The 100 ohm resistor value is measured between the EC 13 SENSE jack and the RETURN jack on the front panel 2 Inductance Jumper For testing to EC...

Page 7: ...4 Figure 2 EC 13 Figures 6 and 6A...

Page 8: ...n the front panel 1 50 OHM jack This is the standard output jack for IEC 60601 1 Table 50 and it is also used for the pulse output for testing in IEC 60601 2 27 This output provides the 50 ohm output...

Page 9: ...6 Figure 3 IEC 60601 2 27 Fig 101...

Page 10: ...7 Figure 4 IEC 60601 2 27 Figure 104...

Page 11: ...8 Figure 5 IEC 60601 2 27 Figure 105...

Page 12: ...eference Connect the MEAS POINT jack to the BLUE electrode on the Pass Fail Reference There is no ground connection on the Pass Fail reference which takes the place of the DUT in EN 60601 Figure 50 Os...

Page 13: ...10...

Page 14: ...6224 Technical Assistance is available from Compliance West USA between the hours of 8 30 AM and 5 00 PM Pacific Time Compliance West USA 2120 Jimmy Durante Blvd Suite 124 Del Mar CA 92014 Phone 858 4...

Reviews: