Compliance Products USA, Inc 
CAL-Fac-001-MP-CDT 240 Rev 2.3   

 

Page 12 of 13 

 
 

9.

 

CBX 1 Second RC Time Constant - Delay Test. 

 

9.1.

 

Line to Ground: Voltage Accuracy. 

 

9.1.1.

 

Edit the test parameters for a Delay test: 

 

Trigger on POS 

 

Delay Test 

 

Delay setting: 1 second 

 

Line to Ground measurement 

 

9.1.2.

 

Connect L-G to 1s time constant terminals “A” and “B” of CAL-BOX and an oscilloscope using a 
10X or 100X probe. 

 

9.1.3.

 

Perform a test and confirm the oscilloscope voltage waveform after 1s of Main relay disconnection 
measurement corresponds to the CDT-240 (within ±2% of reading ±0.8V). 
 

9.1.4.

 

Record on section 7 of CAL-Log-001-CDT 240. 

9.1.4.1.

 

Expected value: Voltage level at 1s. 

Note: 

Compensate for oscilloscope offset. 

9.1.4.2.

 

Tolerance voltage: Waveform value voltage ±2% of reading ±0.8V. 

9.1.4.3.

 

As left voltage: This is the value of the voltage result on the CDT-240. 

 
 

9.2.

 

Line to Neutral: Voltage Accuracy. 

 

9.2.1.

 

Edit the test parameters for a Delay test: 

 

Trigger on POS 

 

Delay Test 

 

Delay setting: 1 second 

 

Line to Neutral measurement 

 

9.2.2.

 

Connect L-N to 1s time constant terminals “A” and “B” of CAL-BOX and an oscilloscope using a 

10X or 100X probe. 

 

9.2.3.

 

Perform  a  test  and  confirm  the  oscilloscope  voltage  waveform  after  1  second  of  Main  relay 

disconnection measurement corresponds to the CDT-240 (within ±2% of reading ±0.8V). 

 

9.2.4.

 

Record on section 7 of CAL-Log-001-CDT 240. 

9.2.4.1.

 

Expected value: Voltage level at 1 second. 

Note:

 Compensate for oscilloscope offset. 

9.2.4.2.

 

Tolerance voltage: Waveform value voltage ±2% of reading ±0.8V. 

9.2.4.3.

 

As left voltage: This is the value of the voltage result on the CDT-240. 

 
 

9.3.

 

Neutral to Ground: Time and Voltage Accuracy. 

 

9.3.1.

 

Edit the test parameters for a Delay test: 

 

Trigger on NEG 

 

Delay Test 

 

Delay setting: 1 second 

 

Neutral to Ground measurement 

 

9.3.2.

 

Reverse the polarity of the EUT voltage by unplugging the EUT power cord, swap the  Line and 

Neutral spade lugs on the power connector, and reconnect EUT power, as it shows it in 

Figure 

7

. 

 

Summary of Contents for CDT-240

Page 1: ...IALS EQUIPMENT NEEDED 5 1 Calibrated DMM Qty 2 DMM1 for voltage RMS below 50V and the DMM2 for voltages RMS 50V or greater 5 2 Calibrated Oscilloscope with 10x or 100x probe 5 3 DC power supply 60V 5...

Page 2: ...document instead of mentioned CAL Log 2 2 Plug in the CDT 240 power cord plug in the EUT power cord to the proper value 90 260V 50 60Hz disconnect all the test leads attached to G L or N outputs disco...

Page 3: ...so that the CDT is not armed The oscilloscope should measure 0V on the BNC connector we expect to get same as Figure 2 adjust P3 in order to set this value Figure 2 Zero output on BNC connector with C...

Page 4: ...be 0 0V because the voltage decayed to zero in 3s Adjust P4 as needed so that the measured voltage is 0 0V turning the potentiometer clockwise makes the reading lower i e more negative Repeat the tes...

Page 5: ...50 0 V Line to Gnd measurement 5 1 3 Press the test button once to ARM the CDT You should see a sine wave signal on the oscilloscope as it shows it in Figure 5 Figure 5 Sine Wave Output on CDT 240 BNC...

Page 6: ...f the voltage is 60VDC for example You can see both TP7 TP6 test points on Figure 6 5 2 3 Record the Neutral to Ground voltage on section 3 N G of CAL Log 001 CDT 240 5 2 3 1 EUT Test Voltage DMM2 Neu...

Page 7: ...EUT voltage is 120V for example To read both meters simultaneously press the HOLD button on both meters at the same time if available 6 1 5 Record both input voltage and L G BNC voltage on section 4 o...

Page 8: ...EUT Test Voltage Expected value 6 2 4 3 Tolerance Expected Value 1 of reading 6 3 Line to Neutral No adjustment verification only 6 3 1 If the EUT connection is reverse connect as normal Figure 7 6 3...

Page 9: ...84 ms for 60 Hz 1 01 ms for 50 Hz 7 2 All measurements are positive peak disconnect POS unless otherwise specified An example of a very accurate disconnect is shown below in Figure 8 Figure 8 Example...

Page 10: ...tion was successful 7 4 5 1 Expected Value Value of previous Positive Sine Wave Peak 7 4 5 2 Tolerance Expected Value 5 0 of reading 7 5 Accuracy at 120V 60Hz input NEG peak 7 5 1 Connect a scope to t...

Page 11: ...he test time result on the CDT 240 8 2 Neutral to Ground Voltage Accuracy 8 2 1 Edit the test parameters for a Voltage test Trigger on NEG Voltage Test Voltage setting 50 0 V Neutral to Ground measure...

Page 12: ...uracy 9 2 1 Edit the test parameters for a Delay test Trigger on POS Delay Test Delay setting 1 second Line to Neutral measurement 9 2 2 Connect L N to 1s time constant terminals A and B of CAL BOX an...

Page 13: ...onfirm the oscilloscope voltage waveform after 1 second of Main relay disconnection measurement corresponds to the CDT 240 within 2 of reading 0 8V 9 3 7 Record on section 7 of CAL Log 001 CDT 240 9 3...

Reviews: