IE-LOG-S0842-01 

 

  9 

 

Warning:

  near  battery  fragile  glass  reed  contact  is  located  –  be  careful  not  to  damage  it.  Be 

careful in battery replacement! 

 

Replacement procedure: 

 

switch off the logger by the PC program or by magnet (if low battery allows) 

 

unscrew four corner screws and remove the lid 

 

remove old battery by pulling the glued tape 

 

insert new battery 

respecting the correct polarity

 (see s and – near battery holder). If 

you  connect  new  battery  up  to  30  s,  all  logger  settings  remain unchanged. In the opposite case 
check by means of PC program all settings, especially the real time clock in logger. 

Attention, 

inserted battery with incorrect polarity causes logger damage!

 

 

put the lid back again and screw four screws. Be careful to have the rubber sealing properly in the 
slot and tighten the screws properly to ensure the water resistance of the instrument. 

 

connect logger to the computer and 

write to it the information on battery replacement 

(menu 

Configuration/Battery  replacement).  This  step  is  necessary  to  evaluate  properly  the  remaining 
battery capacity. 

 

Old battery or logger itself (after its life) is necessary to liquidate ecologically! 
 
 

 

Connection of female ELKA 3008V connector for temperature probe and binary contact 

It  is  possible  to  order  as  an  optional  accessory  connector  for 

connection of temperature probe. Signal cable should be shielded, round 
cross section of minimum diameter of 3.5 mm and maximum diameter of 
5 mm, maximum length of 10 m and as large as possible cross section of 
cords  (stranded  wires).  Outer  diameter  must  be  from  1  to  1.55  mm 
including  insulation.  Connect  connector  to  temperature  signal 
accordingly  to  the  figure  by  means  of  the  manual  included  in  delivery. 
Choose  the  most  suitable  plastic  insert  accordingly  to  the  diameter  of 
stranded  wires  of  the  cable.  The  IP67  protection  is  guaranteed  only  if 
connector is assembled correctly! 

It  is  necessary  to  select  and  connect  cable  for  binary  inputs  with 

respecting  the  same  rules  as  above.  Connect  the  contact  instead  of  the  RTD  probe.  Due  to  high 
impedance of binary inputs it is necessary to avoid places with higher electromagnetic interference 
when installing cables from contacts (switchboards, motors, contactors, frequency converters, etc.) 

 

 

Instrument passed through tests of electromagnetic compatibility (EMC): 

 

Device conforms in accordance with EN 61326-1 these norms: 

 

radiation:  

EN 55011 Class B 

 

immunity:  

EN 61000-4-2 (levels 4/8 kV, Class A) 

 

 

 

EN 61000-4-3 (intensity of electric field 3 V/m, Class A) 

 

 

 

EN 61000-4-4 (levels 1/0.5 kV, Class A) 

 

 

 

EN 61000-4-6 (intensity of electric field 3 V/m, Class A) 

Summary of Contents for S0842

Page 1: ...LOGGER S0842 Three channel temperature logger with one binary input and LCD display Instruction Manual...

Page 2: ...tion is enabled or disabled for each channel individually Binary channel has no alarm mode Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest...

Page 3: ...of internal RTC 200 ppm i e 0 02 17 28 s in 24 h Power Lithium battery 3 6 V size AA Battery life typical data download to PC is carried out about once a week 5 years in continuous on line mode with...

Page 4: ...eck or optionally set the logger real time clock select suitable logging interval it has no influence to binary inputs record select logging mode cyclic or non cyclic set for each channel type of temp...

Page 5: ...x memory could be stored Notice If logger operates as permanently connected to the computer using of magnet start stop is disabled To enable the logger control by magnet is suitable only in cases when...

Page 6: ...icates logger is ON If information on actual measured values is needed it is possible anytime to display reading display by means of the magnet only if communication adapter is not connected permanent...

Page 7: ...ndicated by INIT reading It can be displayed for about 12 s If battery voltage drop occurred since last internal clock setting below critical limit or battery disconnection for longer period than appr...

Page 8: ...ot possible to combine this function with function of switching logger OFF by magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait...

Page 9: ...for temperature probe and binary contact It is possible to order as an optional accessory connector for connection of temperature probe Signal cable should be shielded round cross section of minimum d...

Reviews: