background image

IE-SNC-Pxxxx-09 

 

  3 

Storing  conditions:

  Temperature  from  -30  to  +80  °C,  humidity  from  0  to  100  %RH  without 

condensation 

Material of the case:

  ABS 

Dimensions:

 see dimension drawing 

Weight:

 P0120 approx. 120 g, P01x2 approx. 140 g, P41x1 approx. 125 g 

Liquidation:

 Device itself (after its life) is necessary to liquidate ecologically! 

 

 

Temperature probe P0120   

 

 

Temperature transducers P41x1 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Temperature probe P0102, P0122 and P0132 

 

 

 

 

 

 

 

 
 

 

 

 

 

          

the stem is situated 

 

 

 

 

 

         in the centre of device 

 

 

 

Installation: 

Described devices are supplied from external power source by a cable. The cable is used for 

transmission of measurement temperature to sensing device too. It is recommended to use shielding 
copper cable (e.g. SYKFY) and the shielding connect  to  ground  terminal  of sensing side.  Outside 
diameter of the cable must be from 4 to 8 mm,  maximal cross-section area 2.5 mm

2

 and maximal 

length 1200m. The cable must be located at indoor rooms (probe is not protected for outdoor use). 
The cable should not be led in parallel along power cabling. Safety distance is up to 0.5 m otherwise 
undesirable induction of interference signals can appear.  