8

 

 

IE-COM-D3121-13

 

Operation of the device 

 

 

Switching ON and OFF the device 

 

Switch ON the instrument by pressing ON/OFF key. After switching ON 
the instrument all symbols on the LCD are displayed. If the ON/OFF key 
is  being  held  pressed,  all  LCD  symbols  are  displayed  till  the  key  is 
released.  In  usual  operation  instrument  then  starts  the  measurement 
mode and actual measured values are displayed. 
 
If  a  voltage  drop  below  critical  limit  occurred  since  last  adjustment  of 
internal  real  time  clock  or  battery  disconnection  longer  than 
approximately 1 min, new instrument initialisation is performed (reading 
INIT could be displayed up to 12 s !). 
 
After  initialization is finished instrument starts to display  internal time to 
warn to incorrect internal time. New time adjustment is performed by the 
computer. Press MENU key to switch instrument to actual measurement 
mode. 
 
It is possible to switch OFF the instrument anytime, all instrument setting 
is  kept  saved.  If  automatic  logging  was  active  before  instrument  is 
switched OFF, logging will continue with the same parameters after new 
switching ON.  

 

 
 
 

 

Displaying of actual measured values 

 

In  this  mode  is  instrument  anytime  after  switching ON. It  is  possible  to  enter this  mode from 
other modes by pressing or by repeating pressing of MENU key.  
 
 

Temperature 

is  displayed  on  the  upper  LCD  line  (°C)  and  relative 

humidity in %RH is displayed on the lower LCD line.  
 

 
 
 

 

Press  DISPLAY  key  to  display  other  readings  - 

temperature  in  °C  on 

upper line and dew point temperature on lower line (°C DP).  
 
 

 

 

Summary of Contents for Commeter D3121

Page 1: ...www cometsystem com Instruction manual Commeter D3121 Digital thermometer hygrometer and datalogger...

Page 2: ...values and minimum a maximum memory 9 Functions and settings available from menu 9 Alarm indication and setting 11 LOGGING MODE 12 Logging start from device keyboard 12 Logging start with clearing me...

Page 3: ...Then it is possible to transfer data to a personal computer Temperature is measured by RTD sensor Ni1000 6180ppm Instrument compares measured values of temperature humidity and dew point with two adju...

Page 4: ...approximately 1m s temperature step 20 C Relative humidity Measuring range 0 to 100 RH Resolution 0 1 RH Accuracy 2 5 RH from 5 to 95 RH at 23 C Response time t90 1 min with plastic sensor cover witho...

Page 5: ...umulator In usual operation from adapter accumulator is charged only with small current If accumulator is totally discharged its full charging in instrument takes approximately 100 hours Instrument wi...

Page 6: ...______________________________________________________________________________________________________________________________________________________________ Backlighting turns off automatically 10...

Page 7: ...Not allowed manipulation It is not allowed to touch the sensors under the cover to prevent damage the sensors The sensors under the cover should not be exposed to direct contact with water or other l...

Page 8: ...n is finished instrument starts to display internal time to warn to incorrect internal time New time adjustment is performed by the computer Press MENU key to switch instrument to actual measurement m...

Page 9: ...isplay minimum resp maximum value of other values Pressing MIN MAX or MENU key again to return to default mode Minimum and maximum memory is cleared from menu after confirmation selection CLR see belo...

Page 10: ...s displayed This value illustrates battery condition Display refresh mode is indicated In the FAST mode refreshment is fastest with regular interval approximately 0 7 s In the dynamic refresh mode DYN...

Page 11: ...of the arrow keys Press and hold the arrow key UP to make value increase fast Press and hold the arrow key DOWN to make value decrease fast Release the arrow key and press ENTER to confirm new limit R...

Page 12: ...dicated by a short beep and position of the record in memory is displayed reading MEMO with the record serial number In this mode it is possible to store max 1000 records It is possible to select one...

Page 13: ...indicated with reading LOG on display If memory is filled over 90 symbol LOG is blinking only in automatic logging Logging start with clearing memory and logging mode selection It is possible to clear...

Page 14: ...de is selected after confirmation by ENTER key logging is active In automatic logging mode i e cyclic or non cyclic last applied storing interval is offered to confirm or modify See above procedure Lo...

Page 15: ...protocol of measurement including instrument serial number and a table of logged values with date and time After disconnection of the communication cable reading PC disappears and instrument is in the...

Page 16: ...t for automatic Installator run If Installator is not run automatically depends on setting of operating system find in file browser file CDSetup exe in main directory of the installation CD and click...

Page 17: ...le of measured values occurs Detailed information on file is available from context menu menu occurs after clicking right mouse button on the position of table In context menu function export to text...

Page 18: ...cessary to replace battery with instrument switched OFF otherwise internal time and setting of d REF and AUDI from menu selections and data in memory HOLD MIN and MAX will be lost For the same reason...

Page 19: ...IE COM D3121 13 19 Technical support and service Technical support and service is provided by distributor Contact is included in warranty certificate...

Reviews: