14

 

 

IE-COM-D3121-13

 

 

 

Clearing  procedure  is  in  progress.  The  duration  is  indicated  by 
displaying of already cleared memory in % (digits 0 to 100). 
 
 
 
 
 
After memory is cleared it is possible to select one of three logging 
modes (MODE) by using arrow keys (non cyclic logging is indicated 
by  symbol  n.CYC,  cyclic  logging  by  symbol  CYCL  and  manual 
logging  by  symbol  Man).  If  manual  logging  mode  is  selected,  after 
confirmation  by  ENTER  key  logging  is  active.  In  automatic  logging 
mode (i.e. cyclic or non cyclic) last applied storing interval is offered 
to confirm or modify. See above procedure. 

 
 
 

 
 

 

 

Logging stop from keyboard

 

 

 

Logging  stop  is  enabled  by  confirmation  of  menu  item  STOP  by 
pressing ENTER key. 
 
 
 
 
 
Reading  STOP  NO  is  displayed  to  enable  logging  stop.  Use  arrow 
keys to modify to STOP YES. Logging is still in progress. 
 
 
 
 
 
Logging  is  stopped  after  confirmation  selection  STOP  YES  by 
pressing  ENTER  key.  Instrument  then  returns  to  default  mode  of 
displaying actual values. 
 

 

 

 

Summary of Contents for Commeter D3121

Page 1: ...www cometsystem com Instruction manual Commeter D3121 Digital thermometer hygrometer and datalogger...

Page 2: ...values and minimum a maximum memory 9 Functions and settings available from menu 9 Alarm indication and setting 11 LOGGING MODE 12 Logging start from device keyboard 12 Logging start with clearing me...

Page 3: ...Then it is possible to transfer data to a personal computer Temperature is measured by RTD sensor Ni1000 6180ppm Instrument compares measured values of temperature humidity and dew point with two adju...

Page 4: ...approximately 1m s temperature step 20 C Relative humidity Measuring range 0 to 100 RH Resolution 0 1 RH Accuracy 2 5 RH from 5 to 95 RH at 23 C Response time t90 1 min with plastic sensor cover witho...

Page 5: ...umulator In usual operation from adapter accumulator is charged only with small current If accumulator is totally discharged its full charging in instrument takes approximately 100 hours Instrument wi...

Page 6: ...______________________________________________________________________________________________________________________________________________________________ Backlighting turns off automatically 10...

Page 7: ...Not allowed manipulation It is not allowed to touch the sensors under the cover to prevent damage the sensors The sensors under the cover should not be exposed to direct contact with water or other l...

Page 8: ...n is finished instrument starts to display internal time to warn to incorrect internal time New time adjustment is performed by the computer Press MENU key to switch instrument to actual measurement m...

Page 9: ...isplay minimum resp maximum value of other values Pressing MIN MAX or MENU key again to return to default mode Minimum and maximum memory is cleared from menu after confirmation selection CLR see belo...

Page 10: ...s displayed This value illustrates battery condition Display refresh mode is indicated In the FAST mode refreshment is fastest with regular interval approximately 0 7 s In the dynamic refresh mode DYN...

Page 11: ...of the arrow keys Press and hold the arrow key UP to make value increase fast Press and hold the arrow key DOWN to make value decrease fast Release the arrow key and press ENTER to confirm new limit R...

Page 12: ...dicated by a short beep and position of the record in memory is displayed reading MEMO with the record serial number In this mode it is possible to store max 1000 records It is possible to select one...

Page 13: ...indicated with reading LOG on display If memory is filled over 90 symbol LOG is blinking only in automatic logging Logging start with clearing memory and logging mode selection It is possible to clear...

Page 14: ...de is selected after confirmation by ENTER key logging is active In automatic logging mode i e cyclic or non cyclic last applied storing interval is offered to confirm or modify See above procedure Lo...

Page 15: ...protocol of measurement including instrument serial number and a table of logged values with date and time After disconnection of the communication cable reading PC disappears and instrument is in the...

Page 16: ...t for automatic Installator run If Installator is not run automatically depends on setting of operating system find in file browser file CDSetup exe in main directory of the installation CD and click...

Page 17: ...le of measured values occurs Detailed information on file is available from context menu menu occurs after clicking right mouse button on the position of table In context menu function export to text...

Page 18: ...cessary to replace battery with instrument switched OFF otherwise internal time and setting of d REF and AUDI from menu selections and data in memory HOLD MIN and MAX will be lost For the same reason...

Page 19: ...IE COM D3121 13 19 Technical support and service Technical support and service is provided by distributor Contact is included in warranty certificate...

Reviews: