14
IE-WFS-KAL-WiFi-01
Relative humidity adjustment
. Enter the measured values of the standard
and the device. Click
Save new calibration
button finally:
If the device measured 73.2 %RH before adjustment, then measures
75.3 %RH after it.
Important note: Capacitive humidity sensors usually drifts at high
humidity and not at low humidity values. In most cases, it is sufficient to
perform a one-point adjustment in higher humidity as show at image: