IE-WFS-KAL-WiFi-01
13
Temperature/Humidity Adjustment (W3710, W3711,
W4710, W3721, W3745, W7710)
Important note
: The calibration constants are not stored in the instrument
itself, but inside the probe. Calibration constants at Digi probe cannot be
restored by backup file downloaded from device. Digi probes can be
restored into factory defaults only.
select
Service
– Digi probe adjustment
menu item:
select Input and click
Read from probe
button:
Temperature adjustment
. Is usually not necessary to adjust the
temperature of Digi probe. For adjustment enter the measured values of the
standard and the device. Click
Save new calibration
button finally:
If the device
measured 25.2 °C before adjustment, then after procedure
will measure
25.5 °C.