Instruction manual for use of temperature and RH logger S3541 

 

Logger is designed for measurement and record of ambient temperature, relative humidity and 

two signals 0-5Vdc. Measuring sensors of temperature and humidity are attached to logger. 
Measured values including calculated dew point temperature are displayed on the two-line LCD 
display and are stored in selectable time interval to internal nonvolatile memory. All logger control 
and setting are performed from the PC and password is applicable. It is enabled to switch ON and 
OFF the logger by delivered magnet (this possibility can be in configuration disabled). Start/stop 
magnet enables also to clear the minimum and maximum value memory. Start/stop magnet enables 
also to clear the minimum and maximum value memory. It is also enabled to program automatic 
start in certain day and time (for one month forward).  

Minimum and maximum measured values can be displayed (display switches to actual 

measured values and min/max values automatically). It is also possible to operate logger with 
switched OFF display. Short display of actual measured values is enabled by means of magnet. 

Switched ON logger every 10 seconds (independently on logging interval) updates MIN/MAX 

memory, compares measured values of each quantity with two adjustable limits for each quantity 
and exceeding of limits is indicated on the display (alarm function). Also memory alarm mode is 
selectable, when alarm is indicated permanently till alarm memory reset. Alarm function is enabled 
or disabled for each quantity individually. 

Logging mode can be adjusted as non-cyclic, when logging stops after filling the memory. In 

cyclic mode oldest stored values are overwritten by new. In addition logging mode can be selected 
when logging is active only if measured value is out of adjusted alarm limits. 

Stored values can be transferred from logger memory to the PC by means of communication 

adapter. Communication adapter can be connected to the logger permanently – data logging is not 
interrupted even if data download appears.  

Logger evaluates minimum battery voltage and its drop below allowed limit is indicated on the 

display. At the same time value of remaining battery capacity is available by means of the PC 
program and appears on the logger LCD in % (every time after switch ON). 

 
Technical parameters: 

 

 
Measuring parameters: 
  Ambient temperature (RTD sensor Pt1000/3850ppm): 

 

Measuring range: -30 to +70 °C 

   

 

Resolution: 0.1 °C 

   

 

Accuracy: ± 0.4 °C 

  Relative humidity (reading is temperature compensated at entire temperature range): 

 

Measuring range: 0 to 100 %RH 

   

 

Resolution: 0.1 %RH 

   

 

Accuracy: ± 2.5 %RH from 5 to 95 %RH at 23 °C 

  Dew point (value calculated from temperature and humidity): 

 

   

 

Range: -60 to +70 °C 

   

 

Resolution: 0.1 °C 

   

 

Accuracy: ± 0.5 °C from 30 to 95 %RH 

  Voltage inputs (IN1, IN2): 

Voltage measuring range: 0 to 5 V dc 
Effective resolution: 13 bits 
Accuracy: ±0,2 % FS 
Input resistance: approximately 100k ohms 

Response time with plastic sensor cover (air flow approximately 1 m/s): 
  temperature: t63 < 2 min, t90 < 8 min (temperature step 20 °C) 

 

  IE-LOG-S3541-04 

Summary of Contents for S3541

Page 1: ...LOGGER S3541 Temperature and relative humidity logger with two 0 5V inputs Instruction Manual...

Page 2: ...ctable when alarm is indicated permanently till alarm memory reset Alarm function is enabled or disabled for each quantity individually Logging mode can be adjusted as non cyclic when logging stops af...

Page 3: ...o EN33 2000 3 normal environment accordingly to appendix NM AE1 AN1 AR1 BE1 Operational position negligible Logger installation by self adhesive Dual Lock applied to clean flat surface Not allowed man...

Page 4: ...tching ON logger or disable this option select if record runs permanently or only if alarm is active If alarms are about to be applied set both limits for each measured quantity and enable alarm optio...

Page 5: ...rs should be powered from one source Reversing of polarity does not damage the logger but logger does not measure correctly zero voltage is recorded permanent reversal of poles is not allowed increase...

Page 6: ...d it is possible anytime to display reading display by means of the magnet only if communication adapter is not connected permanently Plug magnet into guide slots from logger front side for approximat...

Page 7: ...isplay switch ON during date and time display all four arrows appears as warning to check or set it again from the computer However all logger functions work without limitation If reading BAT is displ...

Page 8: ...lues will be cleared Battery replacement Low battery is indicated on the display by blinking of reading BAT It can be displayed permanently if battery voltage is too low Replace the battery for new on...

Page 9: ...5 kV class A EN 61000 4 6 only types with external probes intensity of electromagnetic field 3 V m class A APPENDIX Interconnection of logger to input signals by means of optional K0945 adapter with...

Page 10: ...sor is possible to power e g from ac dc adapter connected via coaxial power connector to K0945 adapter Supply voltage and the power of the applied ac dc adapter depend on types of connected sensors 10...

Reviews: