P
P
a
a
g
g
e
e
1
1
5
5
o
o
f
f
3
3
8
8
I
I
N
N
S
S
T
T
R
R
U
U
C
C
T
T
I
I
O
O
N
N
M
M
A
A
N
N
U
U
A
A
L
L
C
C
D
D
N
N
M
M
-
-
s
s
e
e
r
r
i
i
e
e
s
s
C
C
O
O
U
U
P
P
L
L
I
I
N
N
G
G
/
/
D
D
E
E
C
C
O
O
U
U
P
P
L
L
I
I
N
N
G
G
N
N
E
E
T
T
W
W
O
O
R
R
K
K
S
S
(
(
C
C
D
D
N
N
s
s
)
)
SECTI ON 4 -
PRODUCT INSTALLATION
1912 1 E l T o ro R d
●
Si l vera do, Ca lifo rni a 9 26 76
●
(9 49) 459 -96 00
●
co m-p o wer. com
Rev072518
4
4
.
.
2
2
S
S
a
a
f
f
e
e
t
t
y
y
C
C
o
o
n
n
s
s
i
i
d
d
e
e
r
r
a
a
t
t
i
i
o
o
n
n
s
s
It is critical that the CDN be installed in a manner which ensures that EACH of
following conditions are satisfied:
• The metal enclosure of the CDN must be connected to the reference ground
plane. The recommended connection method is via a direct, surface to
surface connection between the CDN base plate and an exposed,
conductive surface of the floor or wall of a shielded enclosure. The CDN base
plate should be bolted to the surface using the mounting holes on the front
and back of the CDN base plate.
• No signal or operating voltages/currents shall exceed those specified in the
product specifications listed in section 3.5.
• The CDN power input port must connected to an appropriate power source
protected by a circuit breaker with a current rating which is greater than or
equal to the rating of the CDN.
Due to the high level of earth leakage current, the device cannot be
connected to any power source protected by a G
round Fault Circuit
Interrupter (GFCI), also called Ground Fault Interrupter (GFI) or Residual
Current Device (RCD).
Failure to comply with any of these points may damage the equipment and/or
pose an electrical hazard.