Capricorn FLX Weather Station
________________________________________________________________________
Columbia Weather Systems, Inc.
Calibrating the Humidity Sensor
The humidity sensor is calibrated at the factory. No field calibration is
required.
The humidity sensor has a ±2% stability over 2 years. Therefore, we
recommend replacing the sensing element every 2 to 4 years.
Calibrating the Rain Gauge Sensor
The rain gauge is calibrated at the factory and does not require any initial
field calibration.
The tipping bucket mechanism is a simple and highly reliable device. The
transmitter must be located in a clear area, away from trees, buildings,
etc. It must also be mounted level. Accurate readings will not be obtained
unless the transmitter is mounted in a level position. The mechanism
must be clean. Any accumulation of foreign material, dust, etc. will alter
the calibration of this unit.
Absolutely accurate calibration can be obtained only with laboratory
equipment, but an approximate field check can be easily made. For field
calibration, a calibration kit is available from the factory.
Calibrating the Solar Radiation Sensor
The solar radiation sensor is calibrated at the factory. No field calibration
is required.