4. Installation Instructions
Testing water quality
The HYDR
IM
L110w G4 is equipped with
a built-in water softening system that must
be adjusted according to the local water
hardness. To read local water hardness,
proceed as follows:
1. The water test kit included with your
HYDR
IM
contains three water
hardness test strips in bags. Take a
water sample from the location where
the machine will be installed.
2. Open one of the bags, remove the
test strip and dip it into the water.
3. Compare th
e color of the strip with
the chart on the back of the bag
Determine the water hardness according
to the c
hart on the water test kit envelope.
4. Power the unit on and select the
Settings key from the main menu.
5. Go to the Setup Menu and select
“Set Regeneration”.
6. Using the up and down arrows, set the
water softener regeneration level
according to the water hardness table
in this section. If your water hardness
falls between two settings, select the
higher setting.
7. Unscrew the water softener container
lid from the bottom right of the chamber
and pour at least 1.0 litres (0.26 gal) of
water
into the water softener container. Then fill
the
salt container to the top (maximum 1 kg/
2.2 lbs.) with salt. Close by screwing the lid
tightly
back into place and run a complete
cycle with no instruments. An improper seal
can lead to corrosion.
*Please note: The water test strip is only accurate up to 250 ppm. If the reading on the test strip exceeds
250 ppm and/or if the location in which the HYDR
IM
is installed has known water quality problems, having
a more detailed and accurate water test done by a test lab is strongly recommended.
Water Hardness Conversion and salt regeneration levels
°dH
US
GPG
PPM
(mg CaCO3 / Litre)
Regen.
1
1.0
18
2
2.1
36
3
3.1
54
Ve
ry
So
ft
4
4.2
71
5
5.2
89
5.6
5.8
100
1
6
6.3
107
6.2
6.4
110
7
7.3
125
8
8.3
143
2
8.4
8.8
150
9
9.4
161
10
10.4
178
10.1
10.5
180
11
11.5
196
3
11.2
11.7
200
11.8
12.3
210
Typically
No Treatment
Necessary
(values from 18-143)
External Treatment
Required
(values >535)
Additional
Water
Treatment
Required
May Require External Treatment
(values from 150-535)
12
12.5
214
13
13.6
232
4
14
14.6
250*
15
15.6
268
16
16.7
286
5
16.8
17.5
300
17
17.7
303
18
18.8
321
19
19.8
339
6
19.6
20.5
350
20
20.9
357
20.2
21.0
360
21
21.9
375
22
22.9
393
22.4
23.4
400
23
24.0
411
24
25.0
428
25
26.1
446
7
25.2
26.3
450
26
27.1
464
27
28.2
482
28
29.2
500
28.6
29.8
510
29
30.2
518
30
31. 3
535
≥
30.3
≥
31.6
≥
540
8
8
NOTE: Values listed below are only valid if a flow
restrictor is installed in HYDRIM L110w G4. Please
verify with your authorized dealer if your HYDRIM
L110w G4 has a flow restrictor installed.