7. Troubleshooting
Page 37
Problem
Possible Causes and Solutions
Instruments are
not clean
•
Do not overload cassettes or instrument baskets.
•
Do not load too many cassettes into the HYDR
IM
.
•
Use the Heavy Duty cycle for cassettes and hinged
instruments.
•
Check filters and wash arms for debris.
•
Increase wash times. See section 4.8.
Touchscreen /
Unit does not come on
•
Ensure that the power button is turned ON.
•
Ensure that the plug is properly inserted into the wall
socket.
•
Check if a fuse or circuit is blown – call service for
assistance.
NOTE: Fuses should NOT be replaced by the operator.
Call an authorized service technician for assistance.
For continued protection against the risk of fire, replace
only with 15A, 250V type F fuses.
The wash arms
do not rotate
•
Remove the wash arms, clean and reinstall them.
•
Check for obstructions in chamber.
The door cannot be
opened
•
The cycle is not complete.
•
The cycle was aborted and the unit has not
completed draining yet.
Operating cycles appear
to be taking too long
to complete
•
The water is not hot enough. Check that the incoming
hot water temperature is in the recommended range.
•
Check that the hot and cold hoses are not reversed.
“No detergent” message
comes on but there is still
a lot of chemical left in the
container
•
Check for a kink in chemical tubing.
•
Check cap for seal.
•
Reinstall chemical solution, see section 6.1
Drying performance is
inadequate
•
Increase the drying time. See section 4.8.
•
Change the HEPA filter.
spare part#: 01-113277S Filter-Air, C61