6.  Care and Maintenance

Tube A

Tube B 
(behind)

Bacteriological filter

Models fitted with a lumens cleaning  
system (LCS) have a biological filter that must 
be replaced every 500 cycles or if dirty. To 
change the biological filter, follow these steps:

1. 

Power the unit OFF and disconnect 
tube A from the bacteriological filter and 
remove the filter from the filter bracket. 

2. 

Note the orientation of the arrow mark on 
the filter before removing. When the filter 
is free of the bracket, carefully discon-
nect tube B from the back of the filter. 

3. 

Before installing the replacement  
bacteriological filter (SciCan order no. 
01-102119S) check that the arrow mark 
on the filter matches the direction of 
the arrow on the bracket. Push the 
right hand filter fitting into tube B.

4. 

Gently press the replacement filter 
into the filter bracket. The arrow mark 
of the filter should be facing out and 
pointing to the right.

5. 

Re-connect tube A to the left hand  
filter fitting. 

6. 

After having replaced the filter, go to 
the "Reset LCS counter" screen in the 
Configuration menu and reset the coun-
ter to zero. (See Section 3.12)

6.4  Filter and wash arm maintenance 

    Filter maintenance

Inspect the coarse and fine filters in the  
bottom of the chamber daily for debris and 
clean if necessary. To clean, remove the  
filter (turn the metal nut at the centre front of 
the filter to release it), rinse under a tap and 
reassemble. Ensure that the filter is firmly 
locked into position when replaced.

3. Remove coarse filter

Summary of Contents for SciCan HYDRIM C61 wd G4

Page 1: ...l rights reserved I SO1588 3 H Y D R I M C 61wd Fully C o m p l i a n t Operator s Manual Benutzerhandbuch USB Manual de uso USB Manuale operatore USB Manuel de l utilisateur USB USB USB Podr cznik ob...

Page 2: ...H info eu scican com Wymagany jest czytnik formatu PDF Aby otrzyma dokument w formie papierowej nale y skontaktowa si z firm SciCan GmbH pod adresem info eu scican com PDF l sare kr vs F r en pappersk...

Page 3: ...escription charts Baskets and accessories Hints and tips for best cleaning results Recommended load configurations Running a cycle Interrupting aborting a cycle Additional user and setup menu function...

Page 4: ...the property of their respective owners Contents of this manual are subject to change without notice to reflect changes and improvements to the HYDRIM product The HYDRIM should only be installed and s...

Page 5: ...than that specified the protection provided by the equipment may be impaired Cleaning solutions may irritate Avoid contact with eyes skin and mouth Never sit stand or lean on the open door The unit ma...

Page 6: ...1 HEPA filter n 1 Dryer exhaust n 1 Door latch n 1 n 1 USB port n 1 LCD touchscreen n 1 Dryer outlet n 1 Water softener n 1 Wash arms top not shown n 1 Coarse filter n 1 Cleaning solution n 1 and dra...

Page 7: ...Page 6 2 Important Information Settings Unlock door Water softener detergent level and door status Network connectivity green when active USB status green when active 2 3 Touchscreen overview Programs...

Page 8: ...een Instructions Instructions Delay Remote Access Prepare for Shipping Set Printer Language Country Date Time Unit No Network Setup Reset Drying Counter Reset LCS Counter Screen Saver Temperature C F...

Page 9: ...g installation If the HYDRIM C61wd G4 is installed in a sterilization center the manufacturer of the sterilization center should allow enough space at the top back and both sides of the unit to facili...

Page 10: ...ersonal protective equipment you use for clinical activities to handle HIP products Always ensure that personal protective equipment if contaminated with HIP Ultra cleaning solution is cleaned or disp...

Page 11: ...ny of these spots or stains a water test may be required to determine the cause The installation of a water treatment system may be required to reduce the amount of dissolved solids in the water and t...

Page 12: ...dd 0 5 kg of water softening salt to the water softener container using the supplied funnel to prevent any salt from spilling into the chamber and close by screwing the lid tightly back into place An...

Page 13: ...ut its policies and programs governing the disposal of electronic devices 3 6 Setting the language The messages displayed by your HYDRIM can be presented in a number of different languages To change t...

Page 14: ...me and select Date Setup 3 From the DATE screen use the keypad to set the date Press EN to save and to return to the Setup menu 4 To change the format in which the date appears return to the Setup men...

Page 15: ...When you have found the amount of time you require press it Press to save and return to the Setup menu 3 14 Adjusting the temperature display 1 2 Scroll to Temperature C F and select 3 Use to choose...

Page 16: ...is 1 Press to save and return to the Setup menu 3 18 Adjusting the screen contrast The touchscreen is calibrated for the lighting condition of most sterilization centers Should you need to adjust the...

Page 17: ...a User PIN From the User PIN screen you can assign up to four PINs To assign a PIN follow these steps 1 2 Scroll to User and select 3 To assign a user PIN select User PIN and use the numeric keypad t...

Page 18: ...s a 10 100Base T Ethernet port located at the back of the unit To connect your HYDRIM to a network using a router follow these steps 1 Connect your network cable to the Ethernet port at the back of th...

Page 19: ...n your network speed and making an initial connection can take longer 3 24 Connecting to a wireless network HYDRIM models offer Wireless capability for connecting units to the G4 network adding flexib...

Page 20: ...the HYDRIM include Stainless steel scalers spatulas and hemostats Stainless steel retractors needle holders skin hooks tissue forceps scissors curettes and other stainless steel instruments free of de...

Page 21: ...e equivalent exposure time to a specified temperature Cycle Prewash Wash Rinse Dry P0 Machine Cleaning Cycle No initial draining 30 C cold 2 10 minutes default 2 minutes N A 30 C cold 2 minutes N A P1...

Page 22: ...vities should be placed so that water can drain freely Large basket 01 113545 Handpiece adapter 01 113625 Hygiene basket 01 113546 Two basket rack 01 113257 Rack for Vertical and Hinged Instruments Re...

Page 23: ...nstruments and ensure that water can flow unrestricted in and out of the tubes Do not place suction tubes horizontally in a basket or in a cassette For instruments containing lumens such as dental han...

Page 24: ...the unit Handle all instruments with care to prevent personal injuries from punctures for example and follow local health and safety regulations to prevent accidents and injuries Using baskets and acc...

Page 25: ...ts solid instruments and open hollow instruments For example Tissue forceps and scissors For example Scalers curettes dental elevators single part mirrors suction tubes etc For example Bone files plug...

Page 26: ...decrease depending on the projected cycle completion 8 The cycle is complete When the cycle is complete the program name and Cycle Complete is displayed and the unit beeps three times Press OK to open...

Page 27: ...TE If the drying phase of the cycle is interrupted the critical parts of the cycle are complete and the message displaying the program name and Cycle Complete Drying interrupted will appear 4 8 Additi...

Page 28: ...ons Users can access a database of pictorial instructions on how to perform routine maintenance such as installing the cleaning solution 4 9 Opening the door in case of a power failure If there is a p...

Page 29: ...on 2 The unit will record up to the last five successful cycles and the last five incomplete cycles If you select a cycle from the list it will display cycle information in a format similar to how it...

Page 30: ...from the web portal 3 When the activity light on your USB storage device stops blinking or the USB icon on the LCD turns from a flashing green to a solid grey remove the USB storage device and transf...

Page 31: ...eaning solution NOTE To avoid spillage replace the solution container only when the message appears and be sure to wear gloves When a red X appears on the water softener detergent icon Press on the ic...

Page 32: ...mark will appear in place of the red X when it is ready for use NOTE The system can also be primed by simply starting a cycle and selecting Detergent Replaced when prompted Place a new bag in the chem...

Page 33: ...ilter and bacteriological filter if fitted Air filter When the message Replace air filter appears the HEPA filter is in need of changing 1 Turn the power off 2 Pull open the ventilation panel below th...

Page 34: ...he arrow mark on the filter matches the direction of the arrow on the bracket Push the right hand filter fitting into tube B 4 Gently press the replacement filter into the filter bracket The arrow mar...

Page 35: ...run a chamber cleaning cycle Press OK to continue the selected wash cycle but remember to run a chamber cleaning cycle at the next available opportunity 6 6 Preparing the unit for shipping If prepari...

Page 36: ...ional guidance documents may require additional testing Please contact your dealer or local medical authorities for details Installation commissioning annual servicing MUST be undertaken by SciCan app...

Page 37: ...dealer for technical service Load carrier check Visually check load carriers trolleys for wear and tear damage and that the runners move freely over their full range Chamber filter check Remove visual...

Page 38: ...ntinued protection against the risk of fire replace only with 15A 250V type F fuses The wash arms do not rotate Remove the wash arms clean and reinstall them Check for obstructions in chamber The door...

Page 39: ...ounter See section 3 12 Air Filter Missing Install Filter Check HEPA filter to ensure it is installed correctly with arrow in correct position Lost USB storage device with all cycle data If you need t...

Page 40: ...from the unit Check that the drain hose is not kinked Check that the filters in the bottom of the chamber are not obstructed Check that the drain outlet is within the specified height noted in the in...

Page 41: ...that water hoses Hot Cold are not reversed Restart cycle CF15 Water Reservoir Overflow Water reservoir overflow switch triggered Restart program Call for service CF16 Ambient Temperature Error Unit i...

Page 42: ...eed too low Check that load is not blocking wash arm Check all holes in wash arms for debris If message persists call for service CF24 Bottom wash arm low RPM Bottom wash arm speed too low Check that...

Page 43: ...invoice identifies the item by serial number and clearly states the date of purchase No other validation is acceptable After one year all SciCan s warranties and other duties with respect to the quali...

Page 44: ...mperature 60 C 140 F Drain 3 4 Drying system Heater 1 kW Electrical connection 200 230VAC 10 single phase 50 Hz 12A 60 Hz 12A Protection class Class I Equipment pollution degree Pollution degree 2 Equ...

Page 45: ...arked as confidential or proprietary at the time of disclosure c Control will mean the possession directly or indirectly of the power to direct or cause the direction of the management and operating p...

Page 46: ...ovide Customer with at least six 6 months advance notice Customer will allow SciCan to use remote access software to aid in the resolution of problems or questions Update fees if any will be billed on...

Page 47: ...release of the SciCan Software Product or b the combination operation or use of the SciCan Software Product with hardware programs or data not furnished or otherwise approved by SciCan if such infrin...

Page 48: ...of Updates or by persons other than SciCan employees or consultants 7 4 DISCLAIMER THE WARRANTIES ABOVE ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED INCLUDING THE IMPL...

Page 49: ...t disclosed or distributed by its employees or consultants in violation of the provisions of this Article 9 The parties agree that the terms and conditions of this Agreement are considered confidentia...

Page 50: ...ing upon written notice to SciCan Customer may assign or otherwise transfer this Agreement to an Affiliate of Customer provided such Affiliate agrees with SciCan to be bound by the terms and condition...

Page 51: ...ht or any other right in the future 10 10 Amendment This Agreement may be amended only by a written document executed by a duly authorized representative of each of the parties 10 11 Exclusive Agreeme...

Page 52: ...mum 0 111W RF Output power Extensible Authentication Protocol While the HYDRIM unit is connected to a WiFi network the security of the connection depends on the configuration of the Wireless infrastru...

Page 53: ...r network it may be worth checking that it s configured for the highest level of security If your network is configured for an older generation of security WEP or WPA Wi Fi Alliance recommends you mov...

Page 54: ...s Setup The HYDRIM unit allows for Wireless or Wired connections but only one type at a time 2 3 1 Selecting between the Wired and Wireless networks 2 3 2 Connecting to a Wireless network 11 WiFi Regu...

Page 55: ...requirements The device has been tested and found to comply with the limits for a Class B digital device pursu ant to FCC Part 15 Subpart B Due to differences in channel allocation if you cannot conn...

Page 56: ...device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can...

Page 57: ...erference received including interference that may cause undesired operation Responsible Party SciCan LTD 1440 Don Mills Road Toronto Ontario Canada M3B 3P9 Telephone 1 800 667 7733 5 2 Industry Canad...

Reviews: