6
5. Sequentially step up to the next larger ParaPost Drill until the
preplanned diameter and depth are achieved
(Fig. 02)
.
6. Use a cylindrical diamond or carbide bur to prepare an
anti-rotational box
(Fig. 03)
.
7. Select the ParaPost post that corresponds to the last drill
used to prepare the post space. Remove head portion from
the post. The head portion is for identification purposes
only. Insert post into post space
(Fig 04)
. Check for occlusal
clearance. Remove the post; shorten if necessary.
8. To prepare for cementation, clean and rinse the canal
(Fig.
05)
and dry with Hygenic Paper Points. Use the dental cement
of your choice according to the manufacturer’s instructions.
9. Use a Lentulo spiral to coat the walls of the post space with
cement
(Fig. 06)
.
10. Coat the post shaft with cement and slowly insert the post
to full depth, allowing excess cement to vent
(Fig.07)
.
11. If necessary, to obtain additional occlusal clearance after the
cement has set, make any adjustments needed to the post
head while holding the post with a hemostat
(Fig. 08)
.
12. After the cement has set, remove any excess before
completing the core and fabricating the final restoration
(Fig. 09)
.
Cleaning and Sterilization
Drills:
Prior to each use, remove debris ultrasonically or by
hand scrubbing with a brush; thoroughly rinse and dry. Place
drills, with a sterilization indicator, into an autoclave bag and
seal. Sterilize in autoclave at 132°C for 12 minutes. Sterilizing
bags should not touch the walls of the autoclave. Once the
sterilization monitor indicates that sterilization was achieved,
place autoclave bag, with the drills inside, in storage until ready
for use. To decontaminate, soak in a 70% isopropyl alcohol
solution for 10 minutes.
Posts:
Prior to each use, place posts, with a sterilization indi-
cator, into an autoclave bag and seal. Sterilize in an autoclave
at 130 - 135 degrees C for 12-15 minutes, following standard
procedures.
Universal Hand Driver:
Prior to each use, remove debris
ultrasonically or by hand scrubbing with a brush. Sterilize by
autoclave or chemical vapor sterilization.
Warning Notice: DO NOT BEND POSTS.
*References and/or literature available upon request.
Summary of Contents for ParaPost
Page 2: ...ParaPost X System Instructions For Use 2...
Page 4: ...ParaPost X System Instructions For Use 4...
Page 88: ...ParaPost X System 88 ParaPostXH 100 102 Hygenic 70 10 130 135 C 12 15...
Page 93: ...ParaPost X System 93 16 Lentulo 17 18 132 C 12 70 10...
Page 100: ...ParaPost XP ParaPost Plus ParaPost One Office Visit Technique 100 01 02 03 04 05 06 07 08 09...
Page 101: ...One Office Visit Technique ParaPost XH 101 10 11 12 13 14 15 16 17 18 19...
Page 102: ...ParaPost XP In Direct Casting Technique 102 20 21 22...