Relay
Action
Effective RELAY-
ON
Set Point
Effective RELAY-
OFF
Set Point
HIGH
S.P.
S.P. – (H.V)
LOW
S.P
S.P. +(H.V.)
S.P. = Relay Set point H.V.= Hysteresis value (Dead Band)
If the relay action is set to
HI
GH, the relay will turn
ON
at (SET
POINT), and will turn
OFF
at (SET POINT-HYSTERISIS).
If the relay action is set to
LO
W, the relay will turn
ON
at (SET
POINT), and will turn
OFF
at (SET POINT + HYSTERISIS).
The figure above shows how the relays react to user set point,
hysteresis and reading.