![Cole Parmer pH 550 Operation Manual Download Page 10](http://html1.mh-extra.com/html/cole-parmer/ph-550/ph-550_operation-manual_2635379010.webp)
10
1.
Temperature
- current temperature of the solution.
2.
pH
- the current degree
of acidity or alkalinity of the
solution with automatic temperature compensation.
B
. CALIBRATION/SETTING MODE
Pressing the [
CAL/VIEW
] key for about two seconds during main
display mode will bring-up the first page of seven pages of the
Calibration/Setting
mode. Pressing [
CAL/VIEW
] key will switch to
the next page until the last page, where pressing [
CAL/VIEW
] again
will return the user to main display mode.
Upper drawing is a simple flowchart showing the path of the
[
CAL/VIEW
] k