12
“
PASSWORD LOCKED
” going to
Calibration/Setting mode
will
not affect the function of the relays.
CAUTION:
If the unit is
not locked
then every time the user enters
the
Calibration/Setting mode
the relays and analog out will be
frozen
.
b.
USER SETTING
page
You will only see this page if the
unit is not password locked. This
page is just a
WARNING
, telling
you that all relays are frozen,
and that you can calibrate and
change the settings.
NOTE:
FROZEN
MEANS ALL THE RELAYS AND THE ANALOG
OUT WILL MAINTAIN THEIR LAST STATE UNTIL THE USER
RETURNS TO MAIN DISPLAY MODE.
c.
pH CALIBRATION
ATC TEMP
. – the current
temperature of the solution.
1.
BUFFER 1
- in this option you
can select which buffer to use for
the standardization calibration.
You can choose
7.00
pH or
6.86
pH by using the [
] and
[
] keys and pressing the [
↵
] key
to save your choice.