5
Safety Cut-outs
The equipment is protected by a miniature circuit breaker and a temperature cut out.
The miniature circuit breakers are located on the rear of the equipment. In the event of a fault, push back to
reset. If the fault continues, please contact your local service representative or the service department of Antylia
Scientific.
If the temperature cut-out is operated allow the equipment to cool completely. Identify and rectify the cause of
cut-out operation before operating. If the temperature cut-out continues to operate, please contact your local
service representative or the service department of Antylia Scientific.
Operation
Before Use
Place the unit on a firm, level surface, ensuring that all four feet on the base of the unit are positioned on the
surface firmly. Avoid installation on a slippery surface or on a surface prone to vibration.
Note:
Button
is not required in this application.
1. Ensure that the power cable is pushed fully into the power supply socket of the wax dispenser.
2. Connect the mains plug to the electrical supply.
3. Pour the desired quantity of wax into the internal tank. The tank has a maximum capacity of 7.5L which is
large enough to melt 6.0kg of pelletized histology wax.
4. Turn ON the wax dispenser.
5. Set the desired temperature. Users are recommended to set the temperature 3-4˚C above the melting point
being used.
a. Press then release it (do not hold down for more than 5 seconds).
b. The display will show alternating with the current set temperature.
c. To change the set temperature press to increase the value or to decrease it. These
keys increase or decrease the value one digit at a time, but if the button is pressed for more than one
second the value increases or decreases rapidly, and after two seconds pressed, the speed increases
even more to allow the desired value to be reached rapidly.
d. Exiting the Set mode is achieved by pressing or automatically if no key is pressed for 15
seconds. After that time the display returns to the normal function mode.