OCCLUSION ADJUSTMENT

(on 77201-60, -62 models only)

Adjust occlusion setting near top of
pump head. 

(Figure #4.)

“4”

for nominal performance;

“5”

for increased pressure/vacuum with

reduced tubing life;

“2”

or 

“3”

for longer life with less

pressure and vacuum.

NOTE:

For optimum performance, turn

knob to 

“1”

after loading. Start drive

and adjust knob until pump primes and
fluid begins to flow.

NOTE:

Occlusion usually does not

need to be readjusted when changing tubing.

MULTI-CHANNEL MOUNTING

L/S EASY-LOAD 

II

Pump Heads can be mounted in tandem (up to four),

depending on the torque capabilities of the drive. 

(See Figure #5.)

Special mounting hardware is required.

See REPLACEMENT PARTS AND ACCESSORIES for mounting hardware.

Figure #5. Multi-channel Mounting

CAUTION:

DO NOT OVER-TIGHTEN SCREWS OR WING NUTS.

NOTE:

The L/S EASY-LOAD 

II

Pump Head can be used in any combination with

other L/S EASY-LOAD Pump Heads. The longer L/S EASY-LOAD 

II

mounting

hardware is required (not supplied).

NOTE:

The tubing on the inside L/S EASY-LOAD 

II

Pump Head(s) can be

changed without removing the outside pump head(s) from the drive.

4

5

INTRODUCTION AND GENERAL DESCRIPTION

This L/S EASY-LOAD 

II

Pump Head, when combined with a MASTERFLEX

®

L/S

®

drive or compatible system, is designed to provide a simple, easy-to-use peristaltic
pump system. The pump head accepts several different tubing sizes for a wide
range of flow rates. The unique over-center cam design and automatic tubing
retention allow quick tubing changes and greatly reduced maintenance time.

Each pump is supplied complete with a 15 in (38-cm) length of tubing, one
mounting hardware package and this manual. A flat-bladed screwdriver or Phillips
is required to complete installation.

DRIVE COMPATIBILITY AND SETUP

The L/S EASY-LOAD 

II

Pump Head is compatible with most MASTERFLEX

®

L/S

®

drives with a standard tang interface. Mount the pump head using the 2 screws
provided. 

(See Figure #1.)

TUBE LOADING

1. Be sure the pump drive is

turned off.

2. Rotate the lever to the 

left to open the pump.

3. Load the correct size

tubing. Center the tubing
between the
retainers.

(Figure #2.)

4. Rotate the lever to the

right to close. 

(Figure #3.)

WARNING:
Stop the drive
when changing
the tubing or its
position in the
rotor mechanism

(the rotor is partially
exposed when the
LOADING LEVER is in
the open position).

NOTE:

For optimum tubing

life, keep tubing straight 
where it enters and exits 
the pump.

LOADING LEVER 

ROTATED TO 

LEFT

TUBING 

CENTERED 

BETWEEN 

RETAINERS

Figure #1. Setup

Figure #2. Open and Load

Figure #3. Close Pump

OCCLUSION SETTING
INDICATOR WINDOW

OCCLUSION 
ADJUSTMENT
KNOB

TUBING 
LOADING
LEVER

Figure #4. Adjust Occlusion