Programmable Option– RS232 Commands and Responses
(Continued)
mode prgm
sets pump in program mode. Response :
number n
sets number of steps in program. n = 1 - 8
step n
sets step number to be programmed. n = 1 - 8
important that step number be set before entering program settings
time xx:xx:xx
sets time duration of step to be programmed. hr:min:sec
travel i (w)
sets direction to infusion ( withdrawal )
rateb nnnnn uuu
sets step beginning rate
where nnnnn is . , 0 to 9
where uuu are units µlm , µlh, mlm, mlh.
Note:
if set rate is out of range then response is NA; rate is set
a zero
ratef nnnnn uuu
sets step finish rate
where nnnnn is . , 0 to 9
where uuu are units µlm, µlh, mlm, mlh.
Note:
if set rate is out of range then response is NA; rate is set
a zero
portout p
sets status of output pins 1 and 6
where p is
HH ( 1=high, 6=high )
HL
( 1=high, 6=low )
LH
( 1=low, 6=high )
LL
( 1=low, 6=low)
pause y or n
sets status of pause
y = active
n = inactive
loop y or n
sets loop status
y = yes
n = no
loopto n
set step number to loop to where n = 1 to 8
loopcnt b
sets number of loops to be repeated
where b = 1 to 100
save
saves step settings
important that each step is saved
done
saves all programmed steps
important that “done” is entered after all steps saved
wait
stops pump (pauses), but can be restarted
continue
restarts pump after “wait” command, program continues
nextstep
causes program to jump to the next step
mode?
query mode. Response PGM
activestep?
queries step running
response: n where n = 1 to 8
Cole-Parmer
®
Dual & Multiple Syringe Pump