background image

21

21

Programmable Option–Features

PARAMETERS WHICH CAN BE PROGRAMMED IN EACH STEP

Time duration

Infuse or withdraw

Start flowrate

End flowrate

TTL output settings

Pause, wait for TTL input actuation

Loops to repeat previous steps

To simplify the programming, previously programmed settings are stored in non-volatile

memory and are displayed when

Program Mode

is selected. Whenever possible, options

are displayed with the “active” option flashing. If flashing, this parameter can be selected

or changed.

The pump can be programmed by first selecting

MODE

on the mainline menu and then

selecting

Program

(PRGM).

DISPLAY AND PROGRAMMING SEQUENCE

After selecting Program Mode, display reads:

Table

Dia

Step

Mode

DIA and TABLE

If the pump was previously used in Program Mode the pump will be initialized in

Program Mode when it is switched on. For convenience, should a syringe change be

required, it is possible to enter a new syringe size, either from the stored Table or DIA,

without leaving the Program Mode.

If a syringe size change is made however, this will change all program values to

defaults and will require reprogramming.

It is possible to review the syringe size in “Dia” or “Table” without changing the pro-

grammed settings.

In “Table” review the settings but select “QUIT”, do not enter a diameter.

In “Dia” the settings will not change if there is no change to the diameter entered.

MODE

Mode selection reverts back to other pump operations.

STEP

Step selection starts the programming sequence.

Cole-Parmer

®

Dual & Multiple Syringe Pump

Summary of Contents for 74900-10

Page 1: ...Dual Multiple Syringe Infusion Pump Users Manual Publication 5607 012 REV B...

Page 2: ...g will be offered for our products which are within the scope of the WEEE Directive Products falling under the scope of the WEEE Directive available for sale after August 13 2005 will be identified wi...

Page 3: ...Setting 12 Rate Setting 12 Power Interruption power up 12 Run Stop 12 Change or Review Volume Setting While Running 13 Mode Selection 13 Infusion Withdrawal Infusion Withdrawal Withdrawal Infusion Con...

Page 4: ...32 Standard Table of Syringe Diameters 33 Japanese Table of Syringe Diameters 34 Standard Minimum and Maximum Flowrates 35 Japanese Minimum and Maximum Flowrates 36 Limited Warranty 37 Cole Parmer Du...

Page 5: ...step 120 secs Linear Travel Rate Min 4 95 x10 4 cm hr Max 12 67 cm min Flowrate range 2 757 l hr to 70 56 ml min 60 ml syringe Nominal linear force 40 lb Dimensions 11x9x5 5 inch 28x23 5x14 cm Weight...

Page 6: ...ECTIONS Make sure all conections are made properly and securely ORIENT EQUIPMENT PROPERLY Do not position the equipment such that it is difficult to reach the disconnecting device OBSERVE ALL TERMINAL...

Page 7: ...syringes arranged by manufacturer material and size Once the syringe is identified in the table the pump automatically enters the appropriate diameter Direct Entry If the syringe used is not included...

Page 8: ...uously The volume is identical in infuse and withdrawal directions Note The displayed menu which prompts the operator for Rate and Volume settings changes with Mode selection For convenience only the...

Page 9: ...the Select key 8 POWER DISRUPTION When power is returned after a temporary power disruption the pump can be pro grammed to resume operation or remain stopped However if a dispense volume is set then...

Page 10: ...Rotate the knob 1 to re engage the drive nut MODEL 210 AND MODEL 230 WITHDRAWAL OPERATION For withdrawal or refill operation the syringe plunger and barrel flange must be secured by additional bracke...

Page 11: ...g large syringes may not require the use of the syringe rest Increase travel is possible if the infusion syringe is retained in the left syringe holder without use of the syringe rest Model 74902 60 h...

Page 12: ...NS A keypad is used for selection of features from the displayed menu and numerical entries KEYPAD These keys move the displayed menu left and right respectively select Used to select the highlighted...

Page 13: ...numerical keys and enter Note If the diameter is changed the volume and rate settings are set to zero VOLUME SETTING Volume can be reached directly from the main menu or will be prompted after mode s...

Page 14: ...Note If the number entered exceeds the maximum flow rate possible then the pump displays the maximum feasible rate To continue enter a rate smaller than the maximum Note To check the maximum possible...

Page 15: ...withdrawal infusion withdrawal withdrawal infusion continuous Note for bi directional modes a volume is required infusion Pump infuses at the set rate and stops automatically when the target volume i...

Page 16: ...e prevented by careful adjustment of the collar position GLASS SYRINGES In the withdrawal mode the retaining bracket on the pusher block clamps on the head of the plunger With some glass syringes the...

Page 17: ...should then be turned off and on to save the settings RS232 Setup The RS232 connections are made through two modular telephone connectors labeled IN and OUT located on the rear panel A single PC can...

Page 18: ...ntrol press select After each command is received and executed by the pump the pump responds with a mes sage and a prompt Commands and Responses Commands are not case sensitive After each command is r...

Page 19: ...sponses Note that mode selection and withdrawal and continuous mode commands are recognized only by the infusion withdrawal models run Starts pump running to present settings returns prompt or If alre...

Page 20: ...l and ml mode I Sets mode to infusion mode w Sets mode to withdrawal mode I w Sets mode to infusion withdrawal mode w I Sets mode to withdrawal infusion mode con Sets mode to continuous Note A dispens...

Page 21: ...also clears all errors A serial error indicates a command that is too long for the input buffer A serial overrun indicates that a command has been sent before the prior command has been processed A st...

Page 22: ...the options available in each STEP The flowrate can be ramped up or down or kept constant for a defined time period for a volume dispense The initial and final flowrate for each period is entered and...

Page 23: ...am Mode display reads Table Dia Step Mode DIA and TABLE If the pump was previously used in Program Mode the pump will be initialized in Program Mode when it is switched on For convenience should a syr...

Page 24: ...ess ENTER to load this time into memory 4 Infusion Withdraw The direction of travel for each step must be selected Initially Infusion direction will be flashing The direction keys and are used to swit...

Page 25: ...on is selected in a step the pump completes the step and pauses at the end of the step The display reads Paused end of n where n is the step number The pump is programmed but stopped waiting to be act...

Page 26: ...ammed c A LOOP within a LOOP It is possible to have a loop running within a loop 9 Saving the Step As there are many options in each step the program gives one more option Redo to make changes before...

Page 27: ...will change all program values to the default settings and will require reprogramming A diameter change causes the pump to stop resets the number of steps to 1 resets the activestep to 1 and all valu...

Page 28: ...case sensitive After each command is received and executed the pump responds with prompt sequence a Query commands carriage return CR line feed text CR line feed 1 or 2 digit address prompt character...

Page 29: ...lm lh mlm mlh Note if set rate is out of range then response is NA rate is set a zero portout p sets status of output pins 1 and 6 where p is HH 1 high 6 high HL 1 high 6 low LH 1 low 6 high LL 1 low...

Page 30: ...h ml m ml h portout queries status of output TTL pins 1 and 6 response HH HL LH LL pause queries whether pause response Y or N where Y is yes N is no loops queries whether loops in program response Sn...

Page 31: ...hing rate 0 1 ml minute loop y Select a loop loopto 1 Program will loop back to step 1 after completeing step 2 loopcnt 1 Will repeat the loop one time save step 3 Ready to program step 3 time 00 00 2...

Page 32: ...unications ground transmit and receive No flow control No handshaking Ground Ground Receive Transmit Ground Ground Receive Transmit OUT IN PC with 9 pin connector PC with 25 pin connector data IN pin...

Page 33: ...s to infuse withdraw mode logic low 0 0 5V max 2ma current sink logic high 2V 5V Fuses The fuses are located in the power entry module on the rear panel The linecord must be removed first to gain acce...

Page 34: ...d lubricated Occasionally a small amount of light machine oil should be applied to the guide rods and a small amount of grease or oil to the leadscrew Solvents of any type should never be used to clea...

Page 35: ...32 57 5 Popper Sons Inc Perfektum glass 0 25 3 45 mm 0 5 3 45 1 4 50 2 8 92 3 8 99 5 11 70 10 14 70 20 19 58 30 22 70 50 29 00 6 Ranfac 2 cc 9 12 mm 5 12 34 10 14 55 20 19 86 30 23 20 50 27 60 7 Scie...

Page 36: ...0 2 5 9 10 5 12 62 10 14 34 20 19 68 30 22 44 50 28 80 100 36 68 6 Nipro 1 ml short 6 61 mm 1 ml long 4 75 3 9 53 5 12 96 10 15 78 20 20 07 30 23 17 50 29 13 7 Hoshi 1 ml sm 4 80 mm 1 ml lg 6 70 2 mm...

Page 37: ...min 3 ml 8 59 mm 0 288 l hr 7360 l min 5 ml 10 3 mm 0 414 l hr 634 ml hr 10 ml 14 57 mm 0 828 l hr 1270 ml hr 20 ml 19 05 mm 1 414 l hr 2171 ml hr 30 ml 21 59 mm 1 817 l hr 2789 ml hr 50 ml 28 9 mm 3...

Page 38: ...87 l hr 2703 l min 2 5 ml 9 1 mm 0 323 l hr 6259 l min 3 ml 9 3 mm 0 337 l hr 8626 l min 5 ml 12 62 mm 0 7 l hr 953 ml hr 10 ml 14 34 mm 0 802 l hr 1230 ml hr 20 ml 19 68 mm 1 51 l hr 2317 ml hr 30 ml...

Page 39: ...ot apply to you There are no implied warranties of merchantability or fitness for a particular use or of any other nature Some atates do not allow this limitation on implied warranty so the above limi...

Page 40: ...Copyright 2010 by Cole Parmer All rights reserved Printed in USA Cole Parmer 625 East Bunker Court Vernon Hills Il 60061 1844 Phone 847 549 7600 Fax 847 247 2929 Web www coleparmer com...

Reviews: