Programmable Option– Running The Program
Run
The Run key starts the program; the displayed time counts down and the direction
arrow flashes.
Hold/Continue
If the Run/Stop key is pressed while running a program, the pump is stopped but gives
an option to end the program, or restart the pump and continue the program to its end.
Program changes when operating
Once a step has commenced no changes are possible in that step. However, while
dispensing changes are permitted to steps still to be executed.
Syringe size changes
If the pump was previously used in Program Mode the pump will be initialized in
Program Mode when it is switched on. For convenience, it is possible to enter a new
syringe size, either from the stored Table or DIA, without leaving the Program Mode.
Note:
If a syringe change is made this will change all program values to the default
settings and will require reprogramming. A diameter change causes the pump to stop;
resets the “number” of steps to 1; resets the “activestep” to 1; and all values will be set
to the initial default settings. The initialization of the new settings takes approximately
two seconds.
Stall Condition
The Fast Forward & Fast Reverse features do not work in Program Mode. Should a stall
occur then go to Infusion Mode where the Fast Forward/Fast Reverse features works,
and use these features to end the stall condition.
By going to the infusion mode the program is still saved in memory.
Changing programs containing loops
Only two loops are permitted in a program and it is necessary to remove from memory
the previously used programs which contain loops. A convenient way to do this is to
input a one step program which does not have a loop. This way, all loops are deleted
from memory and the new program containing loops can be added.
Cole-Parmer
®
Dual & Multiple Syringe Pump