1
1
1
1 7
7
7
7
MECHANICAL
Mech-1 Door not sealing
1/ check physical condition of gasket
2/ assure that gasket is in original location
Mech-2 Motor doesn't move
1/ if shaft spins freely: check connections to motor
and check voltage to motor;
2/ if shaft rubs or is frozen, relieve binding and retest
Mech-3 Motor makes noise
1/ If noise is from the motor, tap the top of motor shaft
with ball peen hammer.
2/ If the sound gets worse, tap the other end of the
shaft Ð avoiding touching the blower wheel.
3/ If there is no change, call Customer Service.
OTHER
Other-1 Controller on at all
times - "locked-up"
1/ turn unit off and on to reset
2/ if cannot change any condition on the front panel,
call Customer Service
Other-2 Controller timer resets
on its own
1/ confirm that power from wall is consistent and
within specs
2/ call Customer Service with serial number
Other-3 Front panel displays
are off
1/ Check for wire damage.
Other-4 Unit or wall
fuse/circuit breaker is
blown
1/ check wall power source
2/ compare current draw and compare to specs on
data plate