Chapter 5
Common Commands
© 2010 Colby Instruments. PDL-100A Operating and Programming Manual Page 39
*ESE
Event Status Enable
COMMAND
The *ESE command sets the Standard Event Status
Enable Register bits. The ESR Register contains a
mask value for the bits to be enabled in the Standard
Event Status Register. A one in the ESR mask will
enable the corresponding bit in the ESR to be
reported, a zero will disable the bit. For example, a
mask value of 255 will enable all bits while a mask
value of 1 will enable only the OPC bit to be set.
Command
Syntax
*ESE
mask_value
Where mask_value is
Bit #
Value Enables
7
128
PON – Power on (not used)
6
64
URQ – User Request (not used)
5
32
CME – Command Error
4
16
EXE – Execution Error
3
8
DDE – Device Dependent Error
2
4
QYE – Query Error
1
2
RQC – Request Control (not used)
0
1
OPC – Operation Complete
Example
*ESE 1
Returns
None
Programming
Example
HP : OUTPUT 705;"*ESE 255"
NI : CWGPIB.Write “*ESE 1” ‘enable OPC bit