Chapter 4
Setting Delay Values
© 2010 Colby Instruments. PDL-100A Operating and Programming Manual Page 18
4.3.4 SETUP Menu – Network Addresses
The current network address, gateway address, network mask, DHCP, and
network port number can be displayed or specified. If no Ethernet cable is
attached or a network session could not be established, the network IP address
reported will be invalid and display 0.0.0.0.
4.3.5 SETUP Menu – RS-232 Terminal Mode or MT-100A Mode
The PDL-100A has one RS-232 port that can be connected to either a MT-100A
Microterminal or to a PC. You must command the PDL-100A to know which
device you have attached else command prompts will not display correctly. The
default mode is MT-100A Mode is ON (i.e. MT-100A Microterminal is
connected the PDL-100A). You can toggle between Command Terminal Mode
(connected to PC) or MT-100A Mode by selecting this option.
You can also turn off or on Command Terminal Mode by sending the MODE
TERM OFF command via GPIB, TCP/IP or RS-232 interface.
MODE TERM OFF specify PC is connected to RS-232 port
MODE TERM ON
specify MT-100A is connected to RS-232 port
Note: To revert to factory default settings for all device addresses and
parameters, set the GPIB address switches located on the backpanel to
ZERO (all OFF) then power-cycle (turn OFF then ON) the PDL-100A
System Unit. Remember to set the GPIB address switches back to a valid
address (1 to 31) after you configure your device settings.