Allow the meter to stabilize for 1 minute after LoZ (Low impedance 300K) function is used.
Guaranteed range of accuracy: range of 5~100%, short circuit is allowed to have remain
readings less than 10 digits.
The AC wave peak factor is up to 3.0 in full value (except 1000V range, this range is 1.5 in full
value).
Non-sine waveform:
Wave peak factor is 1.0 to 2.0, accuracy should be added 1.0%
Wave peak factor is 2.0 to 2.5, accuracy should be added 2.5%
Wave peak factor is 2.5 to 2.0, accuracy should be added 4.0%
Maximum input voltage: 750Vrms
* Range: Measured value = Displayed value - Probe short circuit value
Overload protection: 1000V
3. Resistance measurement
Range
Resolution
Accuracy
600.0Ω*
0.1Ω
±
(1.0%+2)
6.000KΩ
1Ω
±(0.8%+2)
60.00KΩ
10Ω
600.0KΩ
100Ω
6.000MΩ
1KΩ
±(1.2%+3)
60.00MΩ
10KΩ
±
(2.5%+5)