Modular Variable Optical Attenuator
|
VOA
Blade
Coherent Solutions
|
VOA
Blade (V1.4)
23
12.4.2.1 Offset
Selecting ‘Offset’ for the
Attenuation Mode
will allow
you to enter the Input and Output Offset values for
the VOA channel.
The VOA channel attenuation offset will be evaluated
using the summation of the attenuation loss within
the VOABlade (Absolute) and the user specified Input
and Output Offsets within this ‘
VOA Attenuation
Offset (dB)
’ field.
Figure 31 - VOA Channel Offset
12.4.2.2 Absolute
When in ‘
Absolute’ Attenuation Mode
the VOA channel attenuation offset will be evaluated using only the
attenuation loss within the VOABlade.
The minimum value will correspond to the factory set insertion loss of the channels.
12.4.2.3 Relative
When in ‘
Relative’ Attenuation Mode
the VOA channel attenuation offset will be evaluated using only the
attenuation loss across the VOA itself.
The minimum value is set to 0.0 dB.
Figure 32 - VOA Channel Offset Modes