Connect the Breakout Cable
CAUTION
:
To reduce emissions, connect the far end of the Breakout cable
shield to frame ground.
Note
:
l
Make sure that the verifier is not receiving power before you perform
any I/O wiring or adjustments to I/O devices.
l
You can clip unused wires short or use a tie made of non-conductive
material to tie them back.Keep bare wires separated from the +24
VDC wire.
1.
Verify that the 24 VDC power supply is unplugged and not receiving power.
2.
Attach the Breakout cable's +24 VDC and Ground to the corresponding
terminals on the power supply.
CAUTION
:
Never connect voltages other than 24 VDC. Always
observe the polarity shown.
3.
Attach the Breakout cable's M12 connector to the DataMan 475 Verifier's
24 VDC connector.
4.
Restore power to the 24 VDC power supply and turn it on if necessary.