Operation
TO ADJUST CALIBRATION
IMPORTANT: Always obtain three (3) test
results from the grain elevator for the
sample being compared. Average these
three (3) results. Compare this average with
the average of three (3) tests by the
moisture tester.
1. Each grain scale can be individually ad-
justed up to 5.0% by increments of 0.1%
to more closely match the results of an
elevator tester.
2. First select the grain to be adjusted.
3. Test the grain selected, using the sample
T
T
you wish to adjust to. Once valid mois-
ture appears, press the CAL button (A).
NOTE:
Calibration of given grain cannot be performed
unless a valid moisture test has been performed.
4. Once CAL button (A) is pressed, tester will now display the obtained moisture
reading and the current offset applied to that moisture range.
5. Press the UP arrow (B) to raise the adjustment amount or press the DOWN
arrow (C) to lower. The tester will add or subtract up to 5.0% by increments of
0.1% to the current moisture range.
6. After the adjustment amount has been selected, press the CAL button (A) to
return to grain tester mode.
NOTE:
Adjustment can NOT be made to factory calibration
r
to produce moisture readings in a tester that shows
“BELOW LIMIT” or “ABOVE LIMIT” readings.
IMPORTANT:
This tester incorporates Multi-Point Calibration for each grain. Therefore, once a
valid test is taken and an adjustment is made, the adjustment will only affect the
moisture range of the sample tested.
8