Troubleshoot
Nucleus 8 Sound Processor User Guide
87
Problem
Resolution
The safety indicator
light on the battery
charger is flashing
orange
1�
If the indicator light flashes orange
immediately, it could be due to the
wrong battery type connected to the
charger, or the battery has reached its
end of life and needs replacing.
2�
If you are using the Y battery charger,
try the rechargeable battery module
on the other connector of the charger.
3�
The room temperature may be outside
the operating range of the charger.
Try using the charger at a room
temperature of 0°C - +40°C (+32°F -
+104°F). Battery charging will
automatically resume once placed in
the correct temperature range.
4�
Charge timeout is reached, if charging
time becomes excessive after 5 hours.
The battery should not attempt to be
charged again. The battery should no
longer be used and should be
replaced.
5�
The USB port or adaptor may not be
the correct type. Try another USB port
or adaptor.
6�
If using a USB hub, too many devices
may be connected to the hub. Try
removing some devices from the hub.
7�
If using a USB hub, it may not be the
correct type. Use a powered hub.
8�
Try a different or new rechargeable
battery module. Aged batteries may
be due for replacement if the charge
can no longer be attained.