Coachcare B02T User Manual Download Page 8

15

14

Refer to picture-07
The monitor is designed to take measurements and store the measurement 
values in memory for two people using User ID A and User ID B. 
1. Sit comfortably in a chair with your feet flat on the floor.
2. Select your User ID (A or B).
Stretch your arm forward on the desk and keep relaxing, make sure the palm 
of hand is upturned. Make sure your arm is in correct position. Avoid body 
movement. Sit still and do not talk or move during the measurement.
After the cuff has been appropriately positioned on the arm and connected
to the blood pressure monitor, the measurement can begin:

a) Download your clinic’s app to your mobile device and follow in-app 
instructions to set up your blood pressure monitor.
b) Open the app, press START on the device and start measuring!

1 ) Operate via the App on smart phone with Bluetooth

5.4. Measurement Procedure

2) Operate on the device

a) Press the START/STOP button. The pump begins to inflate the cuff. In the 
display, the increasing cuff pressure is continually displayed.
b) After automatically reaching an individual pressure, the pump stops and 
the pressure slowly falls. The cuff pressure is displayed during the 
measurement.
c) When the device has detected your pulse, the heart symbol in the display 
begins to blink.
d) When the measurement has been concluded, the measured systolic and 
diastolic blood pressure values, as well as the pulse will be displayed.
e) The appearance of this      symbol signifies that an irregular heartbeat was 
detected. This indicator is only a caution. It is important that you are relaxed, 
remain still and do not talk during measurements.
f) The measurement results are displayed until you switch the device off. If 
no button is pressed for 60 seconds, the device switches off automatically.
g) Cuff self-checking symbol (     )
The cuff correct symbol (     ) will be displayed if the cuff position is correct, 
otherwise the wrong symbol (     ) will be displayed. Please check the cuff 
again if the wrong symbol (     ) is displayed.
h) Movement error symbol (     )
The Movement Error Symbol (      ) is displayed if you move your body during 
the measurement. Please remove the cuff, and wait 2-3 minutes. Reapply 
the cuff and take another measurement.

Patient Position:
1) Comfortably seated
2) Legs uncrossed
3) Feet flat on the floor
4) Back and arm supported
5) Middle of the CUFF at the level of the right atrium of the heart

NOTE: 

Summary of Contents for B02T

Page 1: ...User manual BEAT Blood Pressure Monitor Model B02T Distributed by CoachCare 97 3rd Avenue New York NY 10003 USA Made in China Model Number B02T ...

Page 2: ...edure for using this device Blood pressure readings obtained by this device should be verified before prescribing or making adjustments to any medications used to control hypertension Under no circumstances should YOU alter the dosages of any drugs prescribed by your doctor This monitor is intended for use by adults only Consult with a physician before using this instrument on a child In cases of i...

Page 3: ... manufacturer for assistance if needed in setting up using or maintaining the device Warning Remove the battery if the equipment is not likely to be used for some time Warning The user must check that the equipment functions safely and see that it is in proper working condition before being used Warning No modification of this equipment is allowed Warning Use of power adapters 1 Adapter input 100 2...

Page 4: ...systolic blood pressure values are between 140 mmHg and 159 mmHg and or the diastolic blood pressure values between 90 mmHg and 99 mmHg consult your physician Regular self checks are necessary If you have blood pressure values that are too low i e systolic values under 105 mmHg and or diastolic values under 60 mmHg consult your physician Even with normal blood pressure values a regular self check ...

Page 5: ...off 10 Volume adjustment 12 Bluetooth function 1 Talking function 3 Cuff self checking function 5 Average value function 7 WHO function 9 External power adapter support 11 Date time display 1 Systolic blood pressure 3 Irregular heartbeat symbol 5 Memory symbol 7 USER A 9 Movement error symbol 11 Average value symbol 13 Cuff self checking function 15 Heartbeat symbol Flashes during measurement 16 WHO ...

Page 6: ... any other trash Battery replacement Configure the system settings after either after loadaing the battery or connecting the monitor 4 2 System Settings A Setting the User Press the SET button and then you can set the A B user by shifting the A B user button B Setting the Year Month Date Time Volume Long press the SET button for more than 3 seconds and then you can configure the settings Setting the...

Page 7: ...le palm upward so the cuff is at the same height as your heart Make sure the tube is not kinked f Remain seated quietly for at least two minutes before you begin the measurement 5 3 Fitting the Cuff Note Comparable blood pressure measurements always require the same conditions Conditions should always be quiet All efforts by the user to support the arm can increase blood pressure Make sure you are in...

Page 8: ...ssure the pump stops and the pressure slowly falls The cuff pressure is displayed during the measurement c When the device has detected your pulse the heart symbol in the display begins to blink d When the measurement has been concluded the measured systolic and diastolic blood pressure values as well as the pulse will be displayed e The appearance of this symbol signifies that an irregular heartbea...

Page 9: ...ed daily or if it suddenly appears more often than usual we recommend the patient to seek medical advice The instrument does not replace a cardiac examination but serves to detect pulse irregularities at an early stage 5 6 Error Indicates The following symbol will appear on the display when measuring abnormal Er 1 SYMBOL No display appears Weak battery or improper placement Sensor abnormal Replace...

Page 10: ...inuing a Measurement Batteries discharged replacements required When the batteries are discharged the battery symbol will flash as soon as the instrument is switched on You cannot take any further measurements and must replace the batteries 5 9 Battery Change Indicator The following symbols may appear in this manual on the BEAT Blood Pressure Monitor B02T or on its accessories Some of the symbols r...

Page 11: ...moving the plug from the socket and reinserting the AC adapter connection Wash hands after each time measurement If one device is used by different patients wash hands before and after each use a Do not expose the device to either extreme temperatures humidity dust or direct sunlight b The cuff contains a sensitive air tight bubble Handle this cuff carefully and avoid all types of stress through twis...

Page 12: ...ature 5 to 40 Humidity 15 to 93 RH Storage And Shipping Conditions Temperature 25 to 70 Humidity 93 RH Atmospheric pressure range 70kPa 106kPa Measuring method Oscillometric Pressure sensor Resistive Measuring range 60 230mmHg SYS 40 130mmHg DIA Pulse 40 to 199 per minute Cuff pressure display range 300mmHg Memory Automatically stores the last 120 measurements for 2 users total 240 Measuring resolu...

Page 13: ... environment NOTE UT is the a c mains voltage prior to application of the test level Guidance and manufacture s declaration electromagnetic immunity Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 10 V m 80 MHz to 2 7 GHz 80 AM at 1kHz 3 Vrms 150 kHz to 80 MHz 3 V RMS outside the ISM band 6 V RMS in the ISM and amateur bands 80 AM at 1kHz Not applicable 10 V m 80 MHz to 2 7 GHz 80 AM at 1kHz ...

Page 14: ...ance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affec...

Page 15: ...28 11 Warranty Card The First Repair The Second Repair Faults Date Repaired by Reasons What is Repaired Faults Date Repaired by Reasons What is Repaired ...