Click “Set Param” to setup the data logging parameters. Set the 
parameters as appropriate for your application. Click the “Manual” 
button to tell the software you want to log data later by pressing the 
REC button on the monitor. Click “OK” to save your settings, then 
disconnect the monitor from your PC. 
 
When you are ready to record, press the REC button on the 
monitor. It will log CO2 levels according to the parameters 
you set in the software. “REC” will flash on the display. The 
flashing frequency is the same as the recording interval you 
set in the “Param” screen. 
 
To save your data log, connect the monitor to your PC, start 
the software and click “Data Download”. You can then view 
the data log as a graph on your PC 
 
NOTE: Every time you click “Set Param” the previously 
logged data in the monitor’s memory is cleared. Therefore, 
you should get in the habit of saving your data files every 
time you create one. 

 

 

Automatic Baseline Calibration 

 

Automatic baseline calibration is useful in areas that are 
regularly unoccupied and CO2 level return to 400ppm every 
few days. ABC is turned off by default. To enable ABC: 

 

Step 1

: Turn the power off. 

Step 2

: While pressing the 

 and

buttons at the same 

time, press the “Power” button.  

 

Step 3

: Release the power button, then release 

 and 

. The screen will show 0PPM (ABC off) or 1PPM (ABC 

on). 

 

Step 4

: Use the 

 or 

buttons to change between ABC 

off (0PPM) and ABC on (1PPM).

 

 

Step 5

: Press the MODE button to confirm. The display 

will flash 30 if ABC is turned off or 31 if ABC is turned on. 

Step 6

: If the next screen shows 1PPM, this indicates 

the change was successful. If it shows 0PPM, the 
change failed. 

Step 7

: Press MODE again to restart the monitor normally. 

 

 

17  www.CO2Meter.com

 

www.CO2Meter.com  18

 

Summary of Contents for XT-10

Page 1: ...tions 2 Reading the Display 3 Using the Monitor 5 Carbon Dioxide Levels and Guidelines 7 Specifications 8 Software Installation 9 Software Application 12 Data Logging 13 Automatic Baseline Calibration 18 Manual Calibration 19 Contact Us 20 Carbon Dioxide Monitor Instruction Manual ...

Page 2: ...ion Package Contents Monitor 110 220VAC to 5V DC Power Adapter USB Cable Instruction Manual Before You Start 1 This monitor contains sensors and microprocessors that use precision electronics It must be kept away from water heat fire combustible liquids and strong electromagnetic interference to prevent damage 2 For accurate results do not block the air sampling inlets Avoid strong or hot winds bl...

Page 3: ...nctions Power Button turns the power ON OFF C F Up Arrow Button change between degrees Fahrenheit and Centigrade When MODE pressed used as up arrow or to increase the value 2 www CO2Meter com www CO2Meter com 2 ...

Page 4: ... arrow or to decrease the value MODE MODE Button hold down for 2 seconds to enter SETTINGS Reading the Display 1 Carbon dioxide display area Battery level indicator PPM Concentration units parts per million CO2 Chemical symbol for carbon dioxide MON month MON DAY MIN SEC REC YEAR RH HOUR 2 3 2 Temperature display area degree Centigrade degree Fahrenheit YEAR year HOUR hour Sound alarm enabled 3 Hu...

Page 5: ...arrow buttons to select the digit to change The selected digit will flash Press the up and down arrows to change it Press the MODE button again for 2 seconds to move to the date and clock values Again use the MODE button to change between values and the up and down arrows to change them When complete press the MODE button to save your changes and return to the normal display The CO2 Alarm Level Da...

Page 6: ...s of concentration and increased likelihood of spreading respiratory viruses like colds etc 5 000ppm OSHA and NIOSH first threshold for safety maximum allowed concentration within a 8 hour working period Specifications Carbon Dioxide Range 0 9999PPM Accuracy 70ppm 3 of reading 0 5 000ppm Resolution 1ppm Response time 1 second Temperature Measurement 10 to 70 14 158 Temperature Resolution 0 1 Tempe...

Page 7: ...ile to your Windows PC Double click the file to start the installation Click Next to begin Accept the default installation folder on your hard drive or click Browse to create a new one Click Next after confirmation and click Next again to start the installation 9 www CO2Meter com www CO2Meter com 10 ...

Page 8: ... cable to connect the computer and the CO2 monitor Click to connect The following prompt indicates successful connection Click to disconnect Setting the Date Clock After connecting to the computer click Set Time to automatically set the clock The monitor will sync with the computer s built in clock 11 www CO2Meter com www CO2Meter com 12 ...

Page 9: ...Click the Immediately button to tell the software you want to log data from the software then click OK to begin data logging While logging REC will flash on the display The flashing frequency is the same as the recording interval you set in the Param screen When you are done data logging click Stop Recording The data log cannot be observed until it is downloaded to your PC To download click Data D...

Page 10: ...a room start recording then come back later and retrieve the monitor and download the data This is useful for measuring air quality in a room over time Start by connecting the monitor to a PC and running the software After connecting click the Set Param button to enter the parameter setting interface as shown in the figure below 15 www CO2Meter com www CO2Meter com 16 ...

Page 11: ...previously logged data in the monitor s memory is cleared Therefore you should get in the habit of saving your data files every time you create one Automatic Baseline Calibration Automatic baseline calibration is useful in areas that are regularly unoccupied and CO2 level return to 400ppm every few days ABC is turned off by default To enable ABC Step 1 Turn the power off Step 2 While pressing the ...

Page 12: ...onitor so that your breath does not influence the manual calibration Manual calibration will take between 5 10 minutes to complete Step 6 After calibration is complete read the display 1 means manual calibration was a success 0 means it failed usually because the CO2 level was not stable during calibration If it fails retry the manual calibration again If it fails repeatedly contact us for support...

Reviews: