6.

 

Man down alarm 

 

Falling by breathing dangerous gases can cause serious injury and even fatality to workers. If the 

Man down alarm function in SAN-10 is set on, SAN-10 can detect falls and send a man down   

alert which will activate the audible and visual alarms and alert other people in the area. 

 

 

The man-down detection uses a three-axis accelerometer to automatically monitor the user’s 

movements. If a sudden fall/impact is followed by a  lack  of  movement  for  6  seconds,  the 

alarm will sound. Turn off the alarm by pressing either of the two buttons. 

 

See Section 2.2 Menu and Configuration Settings, to change settings. 

 

7.

 

Reset Button 

Users can reset the unit by pushing a reset button through a 

hole on back of shell. 

 

 

 

 

 

 

 

 

MAINTENANCE

 

 

Calibration

 

The SAN-10 comes pre-calibrated from the factory. However, the CO2 sensor should be calibrated 
at least once a year, or as described in your company’s safety procedures. You can perform the 

calibration yourself, or you can return it to CO2Meter for factory calibration at a nominal fee. 

 

The temperature and humidity sensors do not require calibration and should remain accurate for 

the life of your unit. 

 

Cleaning and Storage

 

Attention: Do NOT use soap, alcohol, aromatic hydrocarbons, or chlorinated solvents for SAN-10 

cleaning purposes. 

The front cover and case can be cleaned with a mild solution of 

detergent and water. 

Apply sparingly with a soft cloth and allow to dry completely prior to using. 

Summary of Contents for SAN-10

Page 1: ...User Manual R1 0 CO2Meter Inc Personal CO2 Monitor SAN 10 Last Revised November 8 2018 Copyright 2018 CO2Meter Inc all rights reserved ...

Page 2: ...CTION 3 FEATURES 3 MONITOR 3 LCD DISPLAY 4 OPERATION 4 MENU CONFIGURATION 5 Calibration 6 Alarm Threshold 6 MAN DOWN FEATURE 7 MAINTENANCE 7 SPECIFICATIONS 8 BUMP TEST PROCEDURE 9 TROUBLESHOOTING 9 WARRANTY 9 CONTACTUS 9 ...

Page 3: ...ble visual strobe and vibrating alarms Large LCD display Man down alarm leveraging accelerometer technology Rugged design with protective rubber enclosure Heavy duty metal clip Front facing sensorunit Rechargeable Li ion Battery 4 2v 1500mAh 18 hours percharge Black Micro USB cable and Wall USB charger Dual calibration methods Ambient Air 300 600ppm and Nitrogen 0 200ppm Automatic atmospheric pres...

Page 4: ...rn the monitor off or change the C F setting 1 Powerbutton 1 1 When the Monitor is turned off press to turn on theunit 1 2 When the Monitor is turned on press for 3 seconds to turn off theunit When the unit is first turned on it performs 5 seconds countdown for Monitor warm up then enters normal display with current CO2 temperature and humidity readings displayed The monitor starts taking measurem...

Page 5: ...t Ambient Air Calibration 300 600ppm N2 N2 means Nitrogen Calibration User presses thepower button to implement Nitrogen Calibration 0 200ppm H0 5 H0 5T H1 0 H1 5 H3 0 Current Defaults are L 0 5 H1 0 High Alarm Threshold setup User presses thepower button to switch the high alarm threshold H0 5 5000 ppm H0 5A 5000 ppm H1 0 10000 ppm H1 5 15000 ppm and H3 0 30000 ppm L0 5 L0 5A L1 0 L1 5 L3 0 Low A...

Page 6: ...esholds 5 000 ppmCO2 low alarm and 1 CO2 Short TermExposure Limit high alarm The alarm thresholds can be changed dependent on the user to L0 5 5000 ppm L0 5T 5000 ppm TWA L1 0 10000 ppm L1 5 15000 ppm or L3 0 30000 ppm If the measured CO2 value exceeds the define Low Alarm threshold the alarm LED will flash red at 2Hz and the audio alarm de de de de will sound If the CO2 level exceeds the High Ala...

Page 7: ...ngs 7 Reset Button Users can reset the unit by pushing a reset button through a hole on back of shell MAINTENANCE Calibration The SAN 10 comes pre calibrated from the factory However the CO2 sensor should be calibrated at least once a year or as described in your company s safety procedures You can perform the calibration yourself oryou can return it to CO2Meter for factory calibration at anominal...

Page 8: ... Pressure Dependence 0 13 of reading per mmHg Response Time About 2 min for 90 of step change Warm up Time 5 seconds at 22 F IAQ LED Display Green 800ppm Yellow 800 1200ppm Red 1200ppm Measurement interval 2 seconds Temperature Sensor Specifications Temperature Range 50 140 F 10 60 C display Display Resolution 0 1 F 0 1 C Display Options F C switchable Accuracy 0 9 F 0 5 C Response Time 5 30 secon...

Page 9: ... will flash and audible alarm will sound b Alarm 2 High 3 0 30 000 ppm LED will flash and audible alarm will sound TROUBLESHOOTING WARRANTY The monitor comes with a one 1 year warranty starting from the date the monitor was shipped to the buyer For more information visit our website www co2meter com pages terms conditions CONTACT US Support CO2Meter com 386 256 4910 Technical Support US OSHA RECOM...

Reviews: