3.

 

Calibration 

There are two calibration methods:  Ambient Air (300-600ppm) and Nitrogen (0-200ppm). 

By pressing the temperature unit button 

 

for 5 seconds, the monitor will enter CO2 calibration 

mode. While in calibration mode,  press the temperature button  

  

quickly to switch between 

AIR (Ambient Air Calibration) or N2 (Nitrogen Calibration). Press the power 
button 

 

toconfirm. 

 

Zero calibration requires 100% nitrogen (0% CO2) calibration gas, a flow regulator and flexible 

tubing. These can easily be obtained online: 

https://www.co2meter.com/collections/accessories/products/demand-flow-regulator

 

 

The Atmospheric calibration will not require additional materials but must be performed  in 

outdoor (300-600ppm fresh air) during the entire procedure. 

See Section 2.2 Menu and Configuration Settings, to change settings. 

 

4.

 

Alarm Threshold 

There are default thresholds: 5,000 ppm CO2 (low alarm) and 1% CO2 (Short Term Exposure Limit) 

(high alarm). The alarm thresholds can be changed dependent on the user, to L0.5 (5000 ppm), 

L0.5T (5000 ppm TWA), L1.0 (10000 ppm), L1.5 (15000 ppm), or L3.0 (30000 ppm). 

 

If  the measured CO2 value exceeds the define Low Alarm threshold, the alarm LED will flash red  

at 2Hz and the audio  alarm  “de  de  …  de  de  …”  will  sound.  If  the  CO2  level  exceeds  the  

High Alarm threshold, the LED will flash  faster  and  the  audio  alarm  will  increase.  To  alert  

users in a noisy environment,  the monitor  will also vibrate. 

See Section 2.2 Menu and Configuration Settings, to change settings. 

 

 

5.

 

Automatic Atmospheric Pressure Compensation 

The displayed CO2 level is affected by atmospheric pressure or change in altitude. To insure 

maximum accuracy, this monitor has automatic atmospheric pressure compensation for CO2 

concentrations by means of a digital atmospheric pressure sensor integrated in the unit. No 

calculations or maintenance  is required. 

See Section 2.2 Menu and Configuration Settings, to change settings. 

Summary of Contents for SAN-10

Page 1: ...User Manual R1 0 CO2Meter Inc Personal CO2 Monitor SAN 10 Last Revised November 8 2018 Copyright 2018 CO2Meter Inc all rights reserved ...

Page 2: ...CTION 3 FEATURES 3 MONITOR 3 LCD DISPLAY 4 OPERATION 4 MENU CONFIGURATION 5 Calibration 6 Alarm Threshold 6 MAN DOWN FEATURE 7 MAINTENANCE 7 SPECIFICATIONS 8 BUMP TEST PROCEDURE 9 TROUBLESHOOTING 9 WARRANTY 9 CONTACTUS 9 ...

Page 3: ...ble visual strobe and vibrating alarms Large LCD display Man down alarm leveraging accelerometer technology Rugged design with protective rubber enclosure Heavy duty metal clip Front facing sensorunit Rechargeable Li ion Battery 4 2v 1500mAh 18 hours percharge Black Micro USB cable and Wall USB charger Dual calibration methods Ambient Air 300 600ppm and Nitrogen 0 200ppm Automatic atmospheric pres...

Page 4: ...rn the monitor off or change the C F setting 1 Powerbutton 1 1 When the Monitor is turned off press to turn on theunit 1 2 When the Monitor is turned on press for 3 seconds to turn off theunit When the unit is first turned on it performs 5 seconds countdown for Monitor warm up then enters normal display with current CO2 temperature and humidity readings displayed The monitor starts taking measurem...

Page 5: ...t Ambient Air Calibration 300 600ppm N2 N2 means Nitrogen Calibration User presses thepower button to implement Nitrogen Calibration 0 200ppm H0 5 H0 5T H1 0 H1 5 H3 0 Current Defaults are L 0 5 H1 0 High Alarm Threshold setup User presses thepower button to switch the high alarm threshold H0 5 5000 ppm H0 5A 5000 ppm H1 0 10000 ppm H1 5 15000 ppm and H3 0 30000 ppm L0 5 L0 5A L1 0 L1 5 L3 0 Low A...

Page 6: ...esholds 5 000 ppmCO2 low alarm and 1 CO2 Short TermExposure Limit high alarm The alarm thresholds can be changed dependent on the user to L0 5 5000 ppm L0 5T 5000 ppm TWA L1 0 10000 ppm L1 5 15000 ppm or L3 0 30000 ppm If the measured CO2 value exceeds the define Low Alarm threshold the alarm LED will flash red at 2Hz and the audio alarm de de de de will sound If the CO2 level exceeds the High Ala...

Page 7: ...ngs 7 Reset Button Users can reset the unit by pushing a reset button through a hole on back of shell MAINTENANCE Calibration The SAN 10 comes pre calibrated from the factory However the CO2 sensor should be calibrated at least once a year or as described in your company s safety procedures You can perform the calibration yourself oryou can return it to CO2Meter for factory calibration at anominal...

Page 8: ... Pressure Dependence 0 13 of reading per mmHg Response Time About 2 min for 90 of step change Warm up Time 5 seconds at 22 F IAQ LED Display Green 800ppm Yellow 800 1200ppm Red 1200ppm Measurement interval 2 seconds Temperature Sensor Specifications Temperature Range 50 140 F 10 60 C display Display Resolution 0 1 F 0 1 C Display Options F C switchable Accuracy 0 9 F 0 5 C Response Time 5 30 secon...

Page 9: ... will flash and audible alarm will sound b Alarm 2 High 3 0 30 000 ppm LED will flash and audible alarm will sound TROUBLESHOOTING WARRANTY The monitor comes with a one 1 year warranty starting from the date the monitor was shipped to the buyer For more information visit our website www co2meter com pages terms conditions CONTACT US Support CO2Meter com 386 256 4910 Technical Support US OSHA RECOM...

Reviews: