Troubleshooting

 

Guide

 

 

Symptom

 

/

 

Issue

 

Possible

 

Cause

 

/

 

Resolution

 

Meter not 
recognized by 
computer/OS 

Your device’s batteries might be depleted. Replace the batteries 
with fresh ones, or make sure the power supply is connected 
properly (use the included power supply only). 

Device 
doesn’t power 
ON 

Make sure the power supply is connected properly and that 
there is power in the outlet the adapter is connected has the 
appropriate voltage. If using batteries, make sure they are not 
depleted. 

The GasLab

®

 

software 
doesn’t start * 

Your software might be out of date. Update your software by 
either visiting our download webpage at 

http://www.co2meter.com/pages/downloads

 or by selecting 

the “

Check

 

for

 

Updates

” under the 

Help

 menu. Make sure your 

PC meets the minimum requirements. 

Slow 
response 

Check the air flow channels to make sure they are not 
obstructed. 

Reading 

doesn’t 
change 

Make sure the meter is in maximum or minimum mode. 

“Bat” and 
green LED 
keep flashing 

The power supply output voltage is inappropriate or batteries 
depleted. Please use the included power supply only or replace 
batteries with a fresh set. 

*For more troubleshooting tips on GasLab

®

 software, see its manual located at 

www.co2meter.com/pages/downloads

USB Driver Installation Instructions 

To install the appropriate USB port drivers compatible with your meter, follow these steps: 

1.

 

Go to 

http://www.ftdichip.com/Drivers/VCP.htm

 and download the package appropriate for the version of 

Windows installed in your computer. 

2.

 

Move the file you downloaded to a location you can easily access. Make sure you have administrator 
privileges. 

3.

 

Extract the file by right clicking on it and selecting 

extract

 

here.

 

4.

 

Go to your computer’s Device Manager in the 

Control

 

Panel

 

For Windows 8, press the Windows key and 'x' at the same time to bring up the start menu then click on 

Device

 

Manager

 

For Windows 7, open the start menu and type 

Device

 

Manager

 in the search bar. 

5.

 

Find the 

Unrecognized

 

USB

 

Device

 in the list (it usually, but not always, has a yellow triangle icon). 

6.

 

Right click the 

Unrecognized

 

USB

 

Device

 item and select 

Update

 

Driver

 

Software

7.

 

Select 

Browse

 

My

 

Computer

 and point to the folder where you extracted the driver files to (step #3). 

 

This will install the necessary drivers to your computer and allow you to use your meter with GasLab

®

. If you have 

multiple sensors you should only have to perform this procedure once; the operating system will automatically 
find the driver for all the other sensors. 

Summary of Contents for CM-0001

Page 1: ...M 0002 CM 0003 CM 0050 CM 0056 See Table 1 on page 7 for details WARNING Install our complimentary GasLab software before connecting your CO2Meter product s to your computer Failure to do so may affec...

Page 2: ...mplimentary GasLab software before connecting your CO2Meter product s to your computer Failure to do so may affect the ability for GasLab to detect your meter If this happens please follow the instruc...

Page 3: ...INSTRUMENT OVERVIEW 8 LCD DISPLAY 8 QUICK START GUIDE 9 POWERING THE METER 9 OPERATION 9 CONNECTING THE METER TO YOUR PC 9 DATA STORAGE 10 DATA LOGGING 10 SETTINGS 11 DISPLAY SETTINGS 11 PORT SETTINGS...

Page 4: ...community CO2Meter Inc appreciates your business and looks forward to working with you and your team in the future Please take some time to read through this manual in order to become familiar with th...

Page 5: ...Optional Accessories If the meter is to operate in environments where humidity is very high 95 RH an Extreme Moisture Filter CM 0103 can be purchased separately This filter will allow free flow of sa...

Page 6: ...T is not supported on Media Center or Tablet editions Installer will optionally install NET Framework Visit www co2meter com pages downloads to download our complimentary GasLab software to your compu...

Page 7: ...d on user selected time intervals Scientific devices such as this sampling meter require users to have intimate knowledge of the meter s features limitations operation required software and specificat...

Page 8: ...percentage format for all other models Logging ON OFF Percentage of free memory Display Shows CO2 concentration level data logging and free memory space Power Jack Connect included power supply Power...

Page 9: ...iod of time batteries should be removed to prevent damage due to potential battery leakage The POWER switch on the left side of the meter turns the unit ON OFF The LOG switch on the right side of the...

Page 10: ...fer to the GasLab User s Manual for more information Data Logging These models feature an internal memory capable of storing data when not connected to a computer This allows you to simultaneously rea...

Page 11: ...ere you can change the date and time on the meter for more accurate reference on the records The default time of your meter could vary depending on your location Review these settings before you start...

Page 12: ...jolted or otherwise mechanically disturbed the sensor can drift requiring recalibration All calibration procedures follow a single point calibration routine that effectively shifts the zero point of...

Page 13: ...or if readings vary greatly 5 60 and 100 Sampling Data Logger These models do not feature Automatic Background Calibration ABC algorithm due to the CO2 scale they measure To ensure the highest accurac...

Page 14: ...other models General Warm up Time 1 min instant measurements Life Expectancy 15 years Maintenance Interval No maintenance required Data Logging Data Points 60 000 Programmable Interval Data Date time...

Page 15: ...set For more troubleshooting tips on GasLab software see its manual located at www co2meter com pages downloads USB Driver Installation Instructions To install the appropriate USB port drivers compati...

Page 16: ...jected to misuse neglect accident modifications or repairs by you or by a third party No employee or reseller of CO2Meter com s products may alter this warranty verbally or in writing Liability All li...

Page 17: ...7 of 17 Contact Us We are here to help For information or technical support please contact us support co2meter com 386 256 4910 Technical Support 386 872 7665 Sales www co2meter com Address CO2Meter I...