2. NIKI V4 Volumetric Infusion Pump Overview
Specifications
Pumping Mechanism:
Linear peristaltic movement
Flow Rate:
Micro - 0.1 to 99.9ml/hr in 0.1ml increments
Macro - 1 to 999ml/hr in 1ml increments
Total Infusible Volume:
Micro - 0.1 to 2,500ml.
Macro - 1 to 9,999ml.
Total Time Setting
99:59 hours
Accuracy:
± 5%.
KVO rate:
0.1 to 5 ml/hr.
Air Sensor:
Ultrasonic
Maximum Pressure:
0.7 bar or 10 psi Adjustable ( high, normal, low )
Power Supply:
100-240 V, 50/60 Hz. 10w 0.3A max.
Battery:
Nickel-Cadmium 7.2V, 0.9Ah.
Battery Operation at 125 ml/hr: 16 hours
Battery Charging:
Automatic when connected to an AC power source
(12 hours)
Alarms:
Bad System
Low & End Battery
Push Run
Air
Fill Set
Bad set
High Pressure (HI PRES) (down occlusion)
Door Open
End of Program (FIN)
Loc
Dimensions:
138 x 138 x 72mm.
5.4’’ x 5.4’’ x 2.8’’ (l x w x h).
Symbol
Attention (Consult accompanying documents)
Classification
IPX4
CLASS 1
Type CF Equipment
(Degree of protection against electrical shock)
Protected against splashing water
(Degree of protection against liquid ingress)
Housing:
ABS (with fire retarded)