Page 1: ...User Manual CLEMENTS Orbital 360 Benchtop Centrifuge CEN 91705 220 240V 50 60Hz Manual No CEN 91705 017 Issue 1...
Page 2: ...nces St Lidcombe NSW 2141 Australia Phone 61 2 9466 5300 Website www medaust com Safety Thank you for purchasing this Clements Orbital 360 Centrifuge For your safety it is imperative that this unit on...
Page 3: ...Warnings and Cautions 6 Installation 7 Controls 8 Operation 10 Parameter Setting Menu 12 Operation Summary 14 User Programs 16 Rotors 17 Accessories and Spares 19 Cleaning and Maintenance 20 Fuse Rep...
Page 4: ...ology for low maintenance operation Accurate digital control of speed and time for repeatable processing High visibility backlit LCD display with touch control panel Electronic lid interlock to ensure...
Page 5: ...l Protection Class I Noise 65dB A Dimensions 370 W x 250 H x 432 D mm Weight 15kg Environment Indoor 10 to 35 C RH 85 Package Contents Item Description Qty CEN 91705 Cell Cycle 360 Centrifuge 240V 50H...
Page 6: ...be observed when handling potentially biohazardous materials Carriers with aerosol protection are to be used Fluid Spill Do not allow fluids to enter the centrifuge If a spill occurs wipe clean payin...
Page 7: ...be placed on a solid level surface in a well ventilated area The back of the centrifuge must be at least 10cm away from any wall or vertical surface Power Requirements Before connecting the centrifug...
Page 8: ...ss hole Control panel Front View Rear View On Off Switch IEC power inlet Fuse holder Condition LCD Display Shows In Spin Cycle Actual and Set Values of Speed RCF and Time Remaining Setting Mode Set Va...
Page 9: ...spin cycle PULSE Button Spins rotor when held down ENTER Button Enters and saves changed parameters Exits parameter setting mode Selects user program Exits program mode INC Button Increases value of s...
Page 10: ...gain being careful not to bend or apply side load to motor shaft 8 Confirm that rotor rotates smoothly and evenly by hand Carrier Fitting swing out style rotors 1 Fit a full complement of compatible c...
Page 11: ...site positions on the rotor Note Even though the Orbital 360 is designed to absorb considerable out of balance forces running the centrifuge with out of balance samples will cause excessive vibration...
Page 12: ...itted Note It is essential that the rotor setting matches the rotor fitted Otherwise the rotor could be subject to speeds in excess of its rating leading to damage to the rotor 1 Ensure the rotor is s...
Page 13: ...ll auto increment or auto decrement if the INC or DEC button is held down 3 Press the ENTER button to save the setting Set Acceleration Profile 1 Press the MENU button until the acceleration field is...
Page 14: ...d value 2 With no button pressed for 5 seconds unit will time out and revert to previously selected user program Select User Program 1 Press PROG button Press INC or DEC buttons to increment or decrem...
Page 15: ...or set in the user program selected matches the actual rotor fitted Otherwise the rotor could be subject to speeds in excess of its rating leading to damage to the rotor START Spin As per selected Use...
Page 16: ...display counting down to zero The rotor will automatically decelerate to zero using the set deceleration profile When the rotor has stopped the beeper will signal the end of the spin cycle 8 Press th...
Page 17: ...17 CLEMENTS Rotors Angle Rotor CEN 91700 001 24 x 10ml Tubes Max rpm 4000 Max RCF 2850 Rotor Setting 1 Swing Rotor CEN 91700 002 12 x 15ml Carriers Max Speed 5000rpm Max RCF 4100 Rotor Setting 2...
Page 18: ...18 CLEMENTS Rotors Swing Rotor CEN 91700 004 4 x 50ml Carriers Max Speed 4200rpm Max RCF 2760 Rotor Setting 3 Swing Rotor CEN 91700 003 24 x 5ml Carriers Max Speed 5000rpm Max RCF 4100 Rotor Setting 4...
Page 19: ...1700 010 Hinge set for Orbital 360 CEN 91700 011 Electronic Lid Lock for Orbital 360 240V CEN 91700 012 Power Transformer for Orbital 260 360 240V CEN 91700 013 Carrier Set 12 x 15ml for Swing Rotor C...
Page 20: ...rilising agents at their standard cleaning dilution can be used on the bowl Remove cleaning agent residue by wiping with a dry paper towel Solvent based cleaning agents and abrasive cleaning agents sh...
Page 21: ...tton pressed Close lid Adjust Lid Closed limit switch E6 Overspeed condition Incorrect rotor setting Power control PCB fault Motor fault SWITCH OFF Verify correct rotor setting Replace power PCB Repla...
Page 22: ...nnectors Rotor does not spin when START button pressed Lid not closed Close lid Loose connecting cables Check and re seat connectors Power transformer fault Replace power transformer Motor fault Repla...
Page 23: ...23 CLEMENTS 0 1000 2000 3000 4000 0 1000 2000 3000 4000 5000 RCF Graphs 0 20 40 60 80 100 120 140 160 100 200 300 400 500 600 700 800 900 1000 RCF Graphs 5 15 ml 50 ml RCF g rpm 50 ml RCF g rpm...
Page 24: ...24 CLEMENTS Notes...
Page 25: ...25 CLEMENTS Notes...
Page 26: ...rchase Conditions of Warranty 1 The product must be returned to the place of purchase with proof of purchase 2 This warranty is only available to the original purchaser of the product 3 The product mu...
Page 27: ...lacement of the product within the terms of this warranty and the total liability of BMDi TUTA Healthcare for loss or damage of every kind whether arising pursuant to the terms of the sale of the prod...
Page 28: ......