Parameter Setting Menu
Parameter Setting Menu
Press
M
ENU
to select parameter setting mode.
Each press of the
M
ENU
button advances to the next parameter.
Press
E
NTER
to save the parameters and exit the parameter setting
mode.
Parameter setting mode is automatically cancelled after a
5 second timeout with no button pressed.
Set Rotor Fitted
Note: It is essential that the rotor setting matches the rotor
fitted. Otherwise the rotor could be subject to speeds in
excess of its rating leading to damage to the rotor.
1. Ensure the rotor is stopped.
2. Press the
M
ENU
button until the rotor field is highlighted.
3. Press the
I
NC
or
D
EC
buttons until the rotor number matches
the rotor that is fitted to the centrifuge. See table page 15.
4. Press the
E
NTER
button to save the setting.
Speed
Time
[rpm]
[minutes]
Acceleration
Deceleration
Rotor Fitted
Factor 1 - 9
Factor 1 - 9
Rotor 1 - 4
Time
[seconds]