Cr 2021-02-01 vs. 2.3. 

User Manual for VCS-20Cr

 

2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This  manual  is  intended  to  provide  the  user  with  all  necessary  information  for  the 
appropriate,  safe  and  independent  operation  of  Cleamix  VCS-20Cr  device.  Instructions 
contain information on technical features, maintenance, storage and disposal of the device. 

 

Read the operating instructions carefully. In case of uncertainty regarding the interpretation 
of these instructions or if you require  additional information, contact your  local Cleamix 
dealer or

 

[email protected]

 

Store operating instructions carefully with the Cleamix VCS-20Cr device.

 

 

 

For the User

 

Disclaimer

 

Given hydrogen peroxide vapour output and concentration values are indicative examples 
and are intended to help the planning and running of decontamination processes in different 
applications. The type and characteristics of the space and surfaces to be decontaminated 
have an impact on the actual process parameters. Hydrogen peroxide vapour is an effective 
biocide  in  a  broad  temperature  range.  Under  certain  circumstances  it  can  create  a 
condensate  with  water  vapour  unfavourable  for  the  supposedly  dry  decontamination 
process.

 

High humidity rates and/or low temperatures are likely to cause condensation. This can be 
prevented  by  conducting  dry,  heated  air  through  the  space  prior  to  hydrogen  peroxide 
vapour production. The lower humidity permits a higher hydrogen peroxide concentration 
and a dry process. 

 

Cleamix  takes  no  responsibility  for  the  process  parameter  choices  applied  in  individual 
decontamination targets. The responsibility lies always with the user.

 

Cleamix reserves the right to make changes to the operating instructions.

 

 
 

 

VCS-20Cr 

 

Cleamix VCS-20Cr

 is a portable decontamination unit designed for demanding hydrogen 

peroxide bio-decontamination processes. Cleamix VCS-20Cr uses up to 50% aqueous 
hydrogen peroxide solution for generating hydrogen peroxide vapour capable of destroying 
even the most resistant microorganisms, like bacterial spores at room temperatures and 
low concentrations.

 

Hydrogen peroxide vapour has good material compatibility and it is residue free 
decomposing into water and oxygen only. 

 

 

 

Summary of Contents for VCS-20Cr

Page 1: ...User Manual VCS 20Cr Decontamination Unit Compatible for software version 2021 02 01 vs 2 3 ...

Page 2: ...hose connector 10 Catalytic Converter 12 HPP272 Sensor 13 Get started 14 Settings 14 Login 14 Self check 15 Supply 15 Set 15 Circulate 16 Cat 16 Dash 16 Prg 18 Save and Load 18 Delete 19 Enviro 19 Trends 20 Importing of measured data 21 Alarms and notifications 22 Tightvnc viewer installation and configuration guide for controlling VCS 20Cr 23 Ending operation 24 Troubleshooting 25 Appendix I 25 A...

Page 3: ...ive biocide in a broad temperature range Under certain circumstances it can create a condensate with water vapour unfavourable for the supposedly dry decontamination process High humidity rates and or low temperatures are likely to cause condensation This can be prevented by conducting dry heated air through the space prior to hydrogen peroxide vapour production The lower humidity permits a higher...

Page 4: ...nly outdoors or in a well ventilated area P260 Do not breathe mist vapours or spray P280 Wear protective gloves protective clothing eye protection face protection P210 Keep away from heat sparks open flames hot surfaces No smoking P220 Keep Store away from clothing flammable materials combustibles Make sure there is no hydrogen peroxide liquid leakage out to the surroundings Use protective cloves ...

Page 5: ...P310 Immediately call a POISON CENTER or doctor P370 P378 In case of fire Use water for extinction Other information Keep container in a cool place out of direct sunlight Store only in vented containers Do not store on wooden pallets Do not return unused material to its original container Avoid contamination Contamination could cause decomposition and generation of oxygen which may result in high ...

Page 6: ...f a large amount has been ingested In case of skin contact may cause burns erythema blisters or even necrosis Indication of immediate medical attention and special treatment needed if necessary Hydrogen peroxide at these concentrations is a strong oxidant Direct contact with the eye is likely to cause corneal damage especially if not washed immediately Careful opthalmologic evaluation is recommend...

Page 7: ... and generation of oxygen gas which could result in high pressures and possible container rupture Empty drums should be triple rinsed with water before discarding Utensils used for handling hydrogen peroxide should only be made of glass stainless steel aluminium or plastic Pipes and equipment should be passivated before first use Hydrogen peroxide should be stored only in vented containers and tra...

Page 8: ...14126 class 6 Pouring liquid hydrogen peroxide from the 20 litre canister into the 1 litre bottle 1 Safety Make sure that you are wearing the proper safety wear non flammable clothing rubber boots protective gloves and protective goggles Make sure that you have plenty of water available for rinsing in case the HP solution gets spilled It s a good idea to keep an emergency kit and eyewash pack near...

Page 9: ...into the apparatus for evaporation SENSOR Vaisala HPP 272 hydrogen peroxide sensor HYDROGEN PEROXIDE BOTTLE Hydrogen peroxide bottle which is placed to the device POWER CORD Power cord for the VCS devices TABLET Tablet for remote controlling and monitoring CATALYTIC CONVERTER For vapor removal POWER CORD FOR CAT C Power cord for the Catalytic Converter ...

Page 10: ...area during the decontamination cycle 10 Keep the area clear from visitors 11 Use warning signs in addition to upfront informing 12 Position the device on a levelled surface in a central location so that the vapour can spread freely into the space 13 Remove cushion from the case it absorbs hydrogen peroxide Power cord Plug in the power cord by pressing and rotating the socket clockwise to lock to ...

Page 11: ...allation is especially important in close loop systems and should not be neglected Always have a H2O2 detector close by Installing the hose connector Take four 4 of the screws part A and corresponding gasket A and install them according to the image 2 in the in and outlets of the gas Make sure to align the gasket A and part A so that no gaps appear and check that the installation is tight and secu...

Page 12: ... reference Installing hoses to the hose connector Install the supplied hose to the connector with the bridge clamps supplied in the package Image 1 to part B Install the hose as shown in image 5 Confirm that the hoses are tightly secured by gently pulling them If you have not ordered hoses from Cleamix Oy and are using your own Cleamix Oy is not responsible if gas leakage occurs due to non conform...

Page 13: ...e by in case of an invalid installation Dräger handheld detector can be purchased from Cleamix Ltd together with the VCS units or separately When placing the VCS unit in its desired location make sure to align the unit properly and so that it is located in upright position and not in angle Ideal location is on table or other flat object so that the unit is not disturbed or moved by accident If you...

Page 14: ...ser serviceable parts inside the sensor body Cleaning and Maintenance You can clean the H2O2 and humidity sensor body and temperature sensor body and cable by wiping them with a soft lint free cloth moistened with water or mild cleaning agent such as isopropanol Do not wipe the filter wiping the filter may block its pores and or deposit residue on the filter If the filter is heavily contaminated r...

Page 15: ...e and password If you forget your password you can enter using Default User login by pressing GO Change password by selecting the user and adding the current password Then press Change Password Admin level user can Manage Users From login page it is possible to continue to Supply or to Selfcheck Turn the power on the device using the power switch VCS 100Cr devices are equipped with Samsung Galaxy ...

Page 16: ...enerating The amount of H2O2 can be RESET The amount left will move to the field Amount of H2O2 ml The amount of H2O2 value cannot be 0 Supply All the buttons with a picture of hourglass needs to be pushed for 2 seconds Edit the treatment values on Set page Set the desired Treatment time Set the Target PPM level Specify RS value to field RS Limit Set the lowest PPM level where the treatment time w...

Page 17: ...t without circulating the hydrogen peroxide before start Cat CAT AUTOMATIC CONTROL The CAT can be controlled automatically Set the Automatic PPM Range where the CAT will automatically turn ON and OFF It doesn t matter which order the numbers will be entered the device will always start the CAT from the highest value The CAT can also be controlled manually by switching CAT Manual Override ON OFF Da...

Page 18: ...ation process STOP Press to stop the decontamination process SELF CHECK Run the self check process Date and time can be edited by pressing the text on the upper edge Pause the decontamination process by pressing Pause if there is need to change the hydrogen peroxide bottle When the bottle is changed continue the decontamination process by pressing again Pause ...

Page 19: ...re we have area that needs to be decontaminated on regular basis With quick programs we can make the regular basis decontamination sessions faster and more efficient since we don t need to set up the settings every time Prg Save and Load Save recipe with the set treatment values Press Save and Enter the Recipe Name Load saved recipes Select the saved recipe from list and press Load ...

Page 20: ...Cr 2021 02 01 vs 2 3 User Manual for VCS 20Cr 19 Delete Delete recipe by selecting the saved recipe and press Delete Enviro Environment page shows the parameters of the surrounding environment ...

Page 21: ...RY Selects the History mode i e earlier captured data can be viewed and scrolled by left right arrow buttons LEGEND Shows the curve colours and to select which curves are shown Upper graph PPM RS RH Lower graph PPM RH and RS change rate and TEMP RH RS The PPM at upper graph has auto scale the ymax value is increased automatically by steps of 50 If ymin ymax values were changed for manual scale the...

Page 22: ...n be imported as Excel table or as PDF document Pressing Update time the time on the report filename will be updated When generating the reports must wait that the message on the left upper corner has disappeared If new report is chosen too early the created file might be faulty DEL LOG Press to empty the log DEL RPTS Select the reports to delete from list Then press Delete selected GEN RPTS The c...

Page 23: ... working unwell with the unit Alarms can be found on the Alarms page The unit will send notifications if there is something critical that has been overlooked Possible notifications Sensor not connected Supply value is zero Treatment time is zero Target PPM is zero Liquid circulation was not done If the temperature is less than 17 C Warning Temperature is low start treatment anyway The unit will wa...

Page 24: ...ions 256 colors Local Cursor Shape Arrow Press OK and then press Connect This is for Windows only Load the TightVNC installation package from the link below and start the installation https www tightvnc com download 2 8 27 tightvnc 2 8 27 gpl setup 64bit msi Steps Next Accept license Next Choose Custom setup type Do not install Server select Entire feature will be unavailable ...

Page 25: ...atalytic converter The ventilation of the space must be planned case specifically and in accordance with valid regulations Note that porous materials within the decontaminated space may increase the hydrogen peroxide aeration time Switch off the device detach the power cord by lifting the latch and turning the socket counter clockwise until it gets loose Detach the hydrogen peroxide sensor by lift...

Page 26: ...it can t be 0 Hydrogen peroxide bottle is getting empty Check fuses F2 F4 Note If the same fuse blows again the device should be delivered to a country specific authorized distributor or the manufacturer for service Heat plate stops heating In case that the heat plate stops heating the unit must be delivered to a country specific authorized distributor or to the manufacturer The device contains an...

Page 27: ...ve humidity RH need to be taken into account when planning a decontamination process using H2O2 vapour The targeted H2O2 ppm level shall be estimated in accordance with the needed exposure time When RH is very high the amount of H2O2 vapour injected to the space is quite limited but is related to the temperature The colder the space the quicker the saturation point is achieved This fact makes the ...

Page 28: ...ials challenge the optimal distribution and delivery of the bio decontamination agent Typical non condensing decontamination cycle is divided into three phases Once the targeted parameters of a decontamination process have been attained for a sufficient exposure period the space has to be properly aerated It is possible to enter the space safely only when the H2O2 vapour concentration gets below 1...

Page 29: ...Cr 2021 02 01 vs 2 3 User Manual for VCS 20Cr 28 ...